FDA Adverse Event Malfunction Summary report: N

DREAMSTATION AUTO BIPAP

MDR report key: 13625221 · Received February 28, 2022

Report

Report Number
2518422-2022-05109
Event Type
Malfunction
Date Received
February 28, 2022
Date of Event
December 15, 2020
Report Date
April 20, 2023
Manufacturer
RESPIRONICS, INC.
Product Code
BZD
UDI-DI
00606959025691
PMA / PMN Number
K131982
Removal / Correction Number
RES 88058
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS MDR IS BEING SUBMITTED AS PART OF A BATCH SUBMISSION OF COMPLAINTS THAT WERE REASSESSED AS REPORTABLE FOAM DEGRADATION COMPLAINTS; DISCOVERED AS PART OF A RETROSPECTIVE REMEDIATION REVIEW.

Additional Manufacturer Narrative · 0

THE MANUFACTURER PREVIOUSLY SUBMITTED MDR 2518422-2022-05109-1 WAS INCORRECTLY CREATED. PLEASE DISREGARD THE PREVIOUS MDR.

Additional Manufacturer Narrative · 0

THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER PREVIOUSLY RECEIVED INFORMATION ALLEGING VISUALISATION OF PARTICLES RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. REPEATED ATTEMPTS TO HAVE THE DEVICE AND COMPONENTS RETURNED FOR EVALUATION AND INVESTIGATION WERE UNSUCCESSFUL. THE MANUFACTURER BELIEVES THEY WILL BE UNABLE TO GATHER ADDITIONAL INFORMATION. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED. SECTION H6 UPDATED IN THIS REPORT.

Description of Event or Problem · 0

THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO THE BIPAP DEVICE'S SOUND ABATEMENT FOAM. THE DEVICE WAS RETURNED AND PARTICLE CONTAMINATION ON TOP AND BOTTOM ENCLOSURES AND REAR PANEL - BLACK DIRT/DUST WAS OBSERVED DURING DEVICE EVALUATION. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
942835 DREAMSTATION AUTO BIPAP VENTILATOR, NON-CONTINUOUS (RESPIRATOR) BZD RESPIRONICS, INC. DSX700H11C 00606959025691

Patients

Seq Age Sex Outcome Treatment
1 Unknown