ABBOTT LOW VOLTAGE LEAD
Report
- Report Number
- 2017865-2022-04211
- Event Type
- Injury
- Date Received
- February 28, 2022
- Report Date
- February 28, 2022
- Manufacturer
- ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
- Product Code
- DTB
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
HEART RHYTHM SOCIETY TITLE: ELECTRICAL ABNORMALITIES WITH ST. JUDE/ABBOTT PACING LEADS: A SYSTEMATIC REVIEW AND META-ANALYSIS (HEART RHYTHM 2021;18:2061¿2069. DOI:10.1016/J.HRTHM.2021.08.026). AUTHORS: ROSHNI V. KHATIWALA, MD; ELIZABETH MULLINS, MD; DALI FAN, MD; UMA N. SRIVATSA, MD, MS, FHRS; SANKET S. DHRUVA, MD, MHS; ADAM OESTERLE, MD LOW IMPEDANCE A072202, PACING PROBLEM A0712.
IT WAS REPORTED THAT ABBOT PACING LEADS WERE USED AS PART OF A STUDY INVOLVING ABBOTT AND NON ABBOTT LEADS. THIS EVENT IS TO REPORT THE OBSERVATIONS ON THE ABBOTT LEADS USED IN THE STUDY. THE STUDY MEASURED THE RATE AT WHICH ABBOTT LEADS DEMONSTRATED ABNORMALITIES ACROSS 8 PREVIOUS STUDIES. THE ABNORMALITIES TRENDED IN THE STUDY INCLUDED LEAD NOISE, ABRUPT CHANGES IN IMPEDANCE (BOTH HIGH AND LOW), LEAD FRACTURES, INSULATION BREAKS, DECREASES IN SENSING INCLUDING A FAILURE TO SENSE, HIGH CAPTURE THRESHOLDS INCLUDING LOSS OF CAPTURE, INAPPROPRIATE PACING INHIBITION, AND INAPPROPRIATE MODE SWITCHES. ABBOTT PACING LEADS WERE FOUND TO HAVE A 6.1% ABNORMALITY RATE. THE STUDY ALSO REPORTED ON RESOLUTIONS TO THE ABNORMALITIES INCLUDING PROGRAMMING CHANGES AND SURGERY INVOLVING EITHER LEAD REVISION OR EXPLANT. THE STUDY FOUND THAT WHILE EITHER OBSERVATION OR PROGRAMMING CHANGES WERE USUALLY SUFFICIENT, IN 1.0% OF CASES LEAD REVISION OR EXTRACTION WAS PERFORMED. THE STUDY CONCLUDED THAT ABBOTT LEADS ARE ASSOCIATED WITH AN INCREASED RATE OF ABNORMALITIES AND INCREASED RATE OF LEAD REPROGRAMMING AND LEAD REVISION OR EXTRACTION OVER OTHER MANUFACTURERS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 883179 | ABBOTT LOW VOLTAGE LEAD | PERMANENT PACEMAKER ELECTRODE | DTB | ST. JUDE MEDICAL, INC.(CRM-SYLMAR) | LOW VOLTAGE LEAD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |