FDA Adverse Event
Other
Summary report: N
ALARIS SYSTEM PUMP MODULE
MDR report key: 1362047
·
Received January 15, 2009
Report
- Report Number
- 2016493-2009-00008
- Event Type
- Other
- Date Received
- January 15, 2009
- Report Date
- January 9, 2009
- Manufacturer
- CARDINAL HEALTH ALARIS PRODUCTS
- Product Code
- FRN
- PMA / PMN Number
- K012383
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
DATE OF EVENT UNK. A NURSE HAD A CONTINUOUS INFUSION RUNNING AT 100ML/HR, VTBI 1000ML AND AFTER 10 HOURS THERE WAS STILL ABOUT 200ML LEFT IN THE BAG. NO PT INJURY. NO FURTHER DETAILS KNOWN. IV PUMP AND SET WERE NOT SAVED. DISCUSSED POSSIBLE CAUSE OF UNDERINFUSION WITH THE ACCOUNT INCLUDING IMPROPER SET LOADING, ISSUE WITH IMPROPER HEAD HEIGHT AND BACK PRESSURE. TIP SHEET ON PROPER SET LOADING WAS PROVIDED TO HELP RE-EDUCATE THE STAFF. SALES REP MADE A SITE VISIT AND REITERATED PROPER SET LOADING AND IV SET UP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALARIS SYSTEM PUMP MODULE | FRN | CARDINAL HEALTH ALARIS PRODUCTS | 8100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |