FDA Adverse Event Malfunction Summary report: N

BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGE

MDR report key: 13619860 · Received February 28, 2022

Report

Report Number
1920898-2022-00129
Event Type
Malfunction
Date Received
February 28, 2022
Date of Event
February 15, 2022
Report Date
March 17, 2022
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: CUSTOMER RETURNED SEVERAL IMAGES OF 4 SYRINGES WITH 0.3ML GRADUATION MARKINGS BUT NO OTHER IDENTIFICATION. THE SYRINGES HAD THEIR NEEDLE SHIELDS AND HUBS SEPARATE FROM THE BARRELS. THE HUBS HAVE BECOME LODGED INSIDE THE SHIELDS. THERE IS NO DAMAGE TO EITHER THE CONNECTORS AT THE DISTAL TIP OF THE BARREL OR THEIR RESPECTIVE NEEDLE HUBS. NO SIGNS OF USE AND NO OTHER DEFECTS FOUND. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 1123341. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. BASED ON THE IMAGES RECEIVED, BD WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF NEEDLE HUB SEPARATION. CAPA#1630423 WAS INITIATED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 3 BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGES EXPERIENCED THE CANNULA BREAKING OFF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ANIMAL CLINIC REPORTED NEEDLE/SHIELD SEPARATION FROM THE BARREL WHEN REMOVING THE SHIELD.

Description of Event or Problem · 0

IT WAS REPORTED THAT 3 BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGES EXPERIENCED THE CANNULA BREAKING OFF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ANIMAL CLINIC REPORTED NEEDLE/SHIELD SEPARATION FROM THE BARREL WHEN REMOVING THE SHIELD.