BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGE
Report
- Report Number
- 1920898-2022-00129
- Event Type
- Malfunction
- Date Received
- February 28, 2022
- Date of Event
- February 15, 2022
- Report Date
- March 17, 2022
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6. INVESTIGATION: CUSTOMER RETURNED SEVERAL IMAGES OF 4 SYRINGES WITH 0.3ML GRADUATION MARKINGS BUT NO OTHER IDENTIFICATION. THE SYRINGES HAD THEIR NEEDLE SHIELDS AND HUBS SEPARATE FROM THE BARRELS. THE HUBS HAVE BECOME LODGED INSIDE THE SHIELDS. THERE IS NO DAMAGE TO EITHER THE CONNECTORS AT THE DISTAL TIP OF THE BARREL OR THEIR RESPECTIVE NEEDLE HUBS. NO SIGNS OF USE AND NO OTHER DEFECTS FOUND. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 1123341. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. BASED ON THE IMAGES RECEIVED, BD WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF NEEDLE HUB SEPARATION. CAPA#1630423 WAS INITIATED.
IT WAS REPORTED THAT 3 BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGES EXPERIENCED THE CANNULA BREAKING OFF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ANIMAL CLINIC REPORTED NEEDLE/SHIELD SEPARATION FROM THE BARREL WHEN REMOVING THE SHIELD.
IT WAS REPORTED THAT 3 BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGES EXPERIENCED THE CANNULA BREAKING OFF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ANIMAL CLINIC REPORTED NEEDLE/SHIELD SEPARATION FROM THE BARREL WHEN REMOVING THE SHIELD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 685796 | BD ULTRA-FINE¿ 3/10ML INSULIN SYRINGE | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 1123341 |