BD DISCARDIT¿ II SYRINGE
Report
- Report Number
- 3002682307-2022-00067
- Event Type
- Malfunction
- Date Received
- February 28, 2022
- Date of Event
- January 26, 2022
- Report Date
- April 13, 2022
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.9. DEVICE AVAILABLE FOR EVAL?: YES. D.9. RETURNED TO MANUFACTURER ON: 28-MAR-2022. H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309110 AND LOT NUMBER 2110221. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECTS AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. TO AID IN THE INVESTIGATION OF THIS ISSUE, PHYSICAL SAMPLES WERE RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH EXAMINATION OF THE SAMPLES, LEAKAGE COULD BE OBSERVED. IT HAS BEEN DETERMINED THAT THE LEAKAGE MOST LIKELY RESULTED FROM DAMAGE TO THE PLUNGER LIP COMPONENT. THIS TYPE OF DAMAGE MAY OCCUR DURING THE HANDLING OF THE PRODUCT THROUGH THE MANUFACTURING PROCESS OR WITHIN THE PLUNGER ASSEMBLY MACHINE.
IT WAS REPORTED THAT THE BD DISCARDIT¿ II SYRINGE LEAKED WHILE SAMPLING THE SALINE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: "... THE SYRINGE WAS LEAKING WITHOUT THIS PROBLEM BEING DETECTED WHEN THE 0.9% NACL WAS SAMPLED... THE LEAK PROBLEM SEEMS RECURRENT DURING SAMPLING."
IT WAS REPORTED THAT THE BD DISCARDIT¿ II SYRINGE LEAKED WHILE SAMPLING THE SALINE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: "... THE SYRINGE WAS LEAKING WITHOUT THIS PROBLEM BEING DETECTED WHEN THE 0.9% NACL WAS SAMPLED... THE LEAK PROBLEM SEEMS RECURRENT DURING SAMPLING."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 685610 | BD DISCARDIT¿ II SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2110221 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |