PENTAX
Report
- Report Number
- 2518897-2022-00195
- Event Type
- Injury
- Date Received
- February 28, 2022
- Report Date
- June 15, 2023
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FDF
- UDI-DI
- 04961333172085
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS MODEL IS CLASSIFIED AS IMPORT FOR EXPORT AND NOT DISTRIBUTED IN THE UNITED STATES, THEREFORE 510K IS NOT APPLICABLE. WE DO HAVE SIMILAR MODEL EC38-I10L-US AVAILABLE IN THE UNITED STATES WITH A 510K NUMBER K131855. (B)(4). THE CUSTOMER OWNED ENDOSCOPE WAS RECEIVED BY PENTAX MEDICAL FOR EVALUATION ON (B)(6) 2021. THE ENDOSCOPE WAS RETURNED TO PENTAX MEDICAL ON (B)(6) 2021. UNDER WORK ORDER (B)(4) AND THE TECHNICIAN. THE ENDOSCOPE WAS INSPECTED BY PENTAX MEDICAL SERVICE UNDER WORK ORDER (B)(4) AND THE SERVICE TECHNICIAN WAS UNABLE TO CONFIRM THE CUSTOMER COMPLAINT BUT DID DOCUMENT THE FOLLOWING INSPECTION FINDINGS: REMOTE CONTROL BUTTON(1): CRACKED, BENDING RUBBER GLUING: DISCOLORED, NOZZLE GLUING: DIRTY, UP PULLEY WIRE: WORN OUT, LEFT PULLEY WIRE: WORN OUT.THE ENDOSCOPE WAS REPAIRED AND APPROVED BY FINAL QC ON (B)(6) 2022 AND WAS DELIVERED TO THE CUSTOMER. MODEL EC38-I10L, SERIAL NUMBER (B)(4) HAS BEEN ROUTINELY SERVICED AT A PENTAX FACILITY SINCE THE DEVICE WAS PUT INTO SERVICE. ON (B)(6) 2022, A DEVICE HISTORY RECORD(DHR) REVIEW FOR MODEL EC38-I10L, SERIAL NUMBER (B)(6) WAS PERFORMED BY THE MANUFACTURER. THE DHR REVIEW CONFIRMED THE ENDOSCOPE WAS MANUFACTURED IN THE MIYAGI FACILITY ON (B)(6) 2015 UNDER NORMAL CONDITIONS, PASSED ALL REQUIRED INSPECTIONS, AND WAS RELEASED ACCORDINGLY. ALSO, THERE WERE NO REWORKS OR CONCESSIONS AND THE DATES OF APPROVAL FOR SHIPMENT AND ACTUAL DATE SHIPPED WERE CONFIRMED FOR (B)(6) 2015. THE INVESTIGATION IS IN-PROCESS. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED WITH THE NEW INFORMATION. MDR 2518897-2022-00194 IS BEING SUBMITTED FOR THE PENTAX MEDICAL VIDEO COLONOSCOPE MODEL EC38-I10L, SERIAL NUMBER (B)(4) FOR THE (B)(6) 2022 AWARE DATE. MDR IS BEING SUBMITTED FOR THE PENTAX MEDICAL VIDEO COLONOSCOPE MODEL EC38-I10L, SERIAL NUMBER (B)() FOR THE (B)(6) 2021 AWARE DATE.
CORRECTION INFORMATION B4: DATE OF THIS REPORT. B5. REFER TO H10. F7: FOLLOWUP #01. F11: UPDATED DATES. F13: UPDATED DATES. SUMMARY: THE USER FACILITY DID DOCUMENT THAT THE FOLLOW PRE-PROCEDURAL CHECKS AND IFUS FOR DEVICES AND ACCESSORIES. THE DEVICE WAS RETURNED AND EVALUATED. IT WAS CONFIRMED THAT THE WATER JET TUBE WAS CLOGGED. BASED ON THE INVESTIGATION RESULTS, IT WAS CONCLUDED THAT THE MOST PROBABLE CAUSE OF THE CLOGGED WATER JET TUBE WAS INSUFFICIENT REPROCESSING AT THE USER FACILITY. IN ADDITION, THE PENTAX MEDICAL HAS NOT RECEIVED ANY FURTHER INFORMATION FOR THIS EVENT AND THEREFORE, CONSIDERS THIS MEDWATCH REPORT CLOSED.
PENTAX MEDICAL WAS MADE AWARE OF A COMPLAINT THAT OCCURRED IN THE UNITED STATES. ON (B)(6) 2021, THE CUSTOMER INITIALLY REPORTED THAT THE DISTAL END OF THE ENDOSCOPE WAS CRUSHED AND CAUSING MUCOSAL TRAUMA TO THE PATIENT DURING PROCEDURE INVOLVING PENTAX MEDICAL VIDEO COLONOSCOPE MODEL EC38-I10L, SERIAL NUMBER (B)(4). IT REPORTEDLY HAPPENED IN MULTIPLE INSTANCES WITH DIFFERENT PHYSICIANS, PATIENTS, AND PROCEDURE ROOMS. THE USER FACILITY RECEIVED THE ENDOSCOPE BACK FROM PENTAX MEDICAL SERVICE REPAIR AND WHEN USING THE DEVICE ON (B)(6) 2022, THE CUSTOMER REPORTED IT AGAIN CAUSED MUCOSAL TRAUMA WITH BLEEDING TO THE PATIENT DURING THE PROCEDURE. THE USER FACILITY HAS PULLED IT OUT OF USE AND RETURNED THE ENDOSCOPE FOR FURTHER EVALUATION. THE USER FACILITY ALSO MENTIONED RESISTANCE AUXILIARY BIOPSY CHANNEL, AND AFTER DISCUSSING WITH THEIR PENTAX SALES REP, REQUESTED THE DISTAL ENDS AND TUBING BE REPLACED. THIS EVENT MEETS THE REQUIREMENTS FOR FDA REPORTABILITY; HOWEVER SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, MANUFACTURER OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
REFER TO H10.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 905767 | PENTAX | VIDEO COLONOCOPE - I10 STANDARD | FDF | HOYA CORPORATION PENTAX TOKYO OFFICE | EC38-I10L | 04961333172085 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |