FREESTYLE LIBRE 3
Report
- Report Number
- 2954323-2022-06693
- Event Type
- Malfunction
- Date Received
- February 28, 2022
- Date of Event
- February 7, 2022
- Report Date
- March 9, 2022
- Manufacturer
- ABBOTT DIABETES CARE LTD
- Product Code
- QLG
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- 003
Narratives
THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, MODEL NUMBER 72114-01 WHICH HAS A SIMILAR PRODUCT DISTRIBUTED IN THE U.S., MODEL NUMBER 71992-01.
THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, MODEL NUMBER 72114-01 WHICH HAS A SIMILAR PRODUCT DISTRIBUTED IN THE U.S., MODEL NUMBER 71992-01.
REPEATED ATTEMPTS BY ADC TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL AND/OR THE CUSTOMER DISCARDED THE PRODUCT. NO PRODUCT HAS BEEN RETURNED. EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. DHRS FOR THE FS LIBRE SENSOR AND FS LIBRE SENSOR KIT WERE REVIEWED AND THE DHRS SHOWED THE FS LIBRE SENSOR AND SENSOR KIT PASSED ALL TESTS PRIOR TO RELEASE. IF THE PRODUCT IS RETURNED, A PHYSICAL INVESTIGATION WILL BE PERFORMED AND A FOLLOW-UP REPORT SUBMITTED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 800236 | FREESTYLE LIBRE 3 | CONTINUOUS GLUCOSE MONITORING SYSTEM | QLG | ABBOTT DIABETES CARE LTD | 72114-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |