FDA Adverse Event Other Summary report: N

VAPOR PHASE PLUS HUMIDIFIER

MDR report key: 136126 · Received December 2, 1997

Report

Report Number
2020364-1997-00024
Event Type
Other
Date Received
December 2, 1997
Date of Event
November 4, 1997
Report Date
December 2, 1997
Manufacturer
SIMS/RESPIRATORY SUPPORT PRODUCTS, INC.
Product Code
BTT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

DURING TREATMENT OF A 75 Y/O MALE PT (WEIGHT UNREPORTED), THE VAPOR PHASE PLUS HUMIDIFIER, REPORTEDLY, OVERHEATED. THE TRANSFER CHAMBER MELTED. IT WAS OBSERVED THAT THREE QUADRANTS OF THE TRANSFER CHAMBER MATERIAL "BURNED". THE HUMIDIFIER WAS RAN DRY. REPORTEDLY, THE DEVICE DID NOT SHUT OFF WHEN THE OVERHEATING OCCURRED; HOWEVER, THE INTERNAL FUSE DID BLOW. THE HUMIDIFIER WAS CONNECTED TO A MODEL 7299 PURITAN BENNETT VENTILATOR; AIR MIX WAS 30%; FLOW RATES ARE UNREPORTED. AN IN-LINE THERMOMETER WAS USED TO MONITOR THE PT'S TEMPERATURE WHEN REGULAR VENTILATOR CHECKS WERE MADE. IT IS UNREPORTED IT THE PT IS TRACHED. A PULMONARY SPECIALIST WAS CONSULTED. THERE WAS NO EFFECT ON THE PT. THE PT IS BEING TAKEN OFF THE VENTILATOR TODAY 11/5/97. THE HOSPITAL RECEIVED ON UPDATED MANUAL NOVEMBER 1995.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VAPOR PHASE PLUS HUMIDIFIER HUMIDIFIER BTT SIMS/RESPIRATORY SUPPORT PRODUCTS, INC. 009300 NA

Patients

Seq Age Sex Outcome Treatment
1 75 YR Other