FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ SYRINGE WITH ATTACHED NEEDLE 18 G X 1 1/2 IN., STERILE

MDR report key: 13598674 · Received February 25, 2022

Report

Report Number
1213809-2022-00100
Event Type
Malfunction
Date Received
February 25, 2022
Date of Event
February 11, 2022
Report Date
March 11, 2022
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903095804
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: SINCE NO SAMPLES DISPLAYING THE REPORTED CONDITION WERE RECEIVED A POTENTIAL ROOT CAUSE COULD NOT BE DEFINED AND CORRECTIVE ACTIONS ARE NOT NECESSARY. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD LUER-LOK¿ SYRINGE WITH ATTACHED NEEDLE 18 G X 1 1/2 IN., STERILE FOREIGN MATTER WAS FOUND IN THE BARREL. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO:309580 BATCH NO: 1081263.   IT WAS REPORTED BY THE CONSUMER OILY SUBSTANCE IN THE BARREL.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD LUER-LOK¿ SYRINGE WITH ATTACHED NEEDLE 18 G X 1 1/2 IN., STERILE FOREIGN MATTER WAS FOUND IN THE BARREL. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO:309580 BATCH NO: 1081263.   IT WAS REPORTED BY THE CONSUMER OILY SUBSTANCE IN THE BARREL.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD LUER-LOK¿ SYRINGE WITH ATTACHED NEEDLE 18 G X 1 1/2 IN., STERILE FOREIGN MATTER WAS FOUND IN THE BARREL. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO:309580. BATCH NO: 1081263.   IT WAS REPORTED BY THE CONSUMER OILY SUBSTANCE IN THE BARREL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1194908 BD LUER-LOK¿ SYRINGE WITH ATTACHED NEEDLE 18 G X 1 1/2 IN., STERILE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 309580 1081263 30382903095804

Patients

Seq Age Sex Outcome Treatment
1 Unknown