S-CATH ESOPHAGEAL TEMPERATURE PROBE
Report
- Report Number
- 3009437315-2022-00002
- Event Type
- Injury
- Date Received
- February 24, 2022
- Date of Event
- January 26, 2022
- Report Date
- February 24, 2022
- Manufacturer
- CIRCA SCIENTIFIC, INC.
- Product Code
- FLL
- UDI-DI
- 10860237000214
- PMA / PMN Number
- K112376
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
PROBE WAS DISCARDED BY THE HOSPITAL AFTER USE AND THEREFORE NO VISUAL INSPECTION OF THE COMPLAINT PROBE COULD BE PERFORMED. PERFORMED A REVIEW OF THE DEVICE HISTORY RECORDS, LABELING AND EVALUATION OF A PROBE FROM RELEASED STOCK. BASED ON EVALUATION OF SAMPLE FROM RELEASED STOCK AND ENGINEERING CHANGE REVIEW, THERE IS NO PRODUCT CHANGE OR DEFECT THAT WOULD ATTRIBUTE TO THE BLEEDING. HOSPITAL COMMUNICATED EGD WAS UNREVEALING WITH NO CLEAR SOURCE OF BLEEDING, MOST LIKELY OROPHARYNGEAL INJURY WITH BLOOD INTO ESOPHAGUS. ESOSURE ESOPHAGEAL RETRACTOR WITH AN OG TUBE WAS ALSO USED DURING THE PROCEDURE.
DURING PROBE INSERTION THEY ENCOUNTERED RESISTANCE. DOCTOR ALSO USES ESOSURE WITH AN OG TUBE. OG TUBE WAS INSERTED BEFORE CIRCA PROBE. TOWARD END OF CASE BLEEDING WAS NOTED. ENT WAS CALLED AND DIDN'T NOTE ANY DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2401020 | S-CATH ESOPHAGEAL TEMPERATURE PROBE | ESOPHAGEAL TEMPERATURE PROBE | FLL | CIRCA SCIENTIFIC, INC. | CS-2001 | 17394-01 | 10860237000214 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Hospitalization | ESOSURE ESOPHAGEAL RETRACTOROG TUBE |