FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 SELF-TEST

MDR report key: 13595276 · Received February 24, 2022

Report

Report Number
1221359-2022-01058
Event Type
Malfunction
Date Received
February 24, 2022
Date of Event
February 2, 2022
Report Date
June 20, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE INVESTIGATION CONCLUSION. PLEASE SEE UPDATES: D4, G3, G6, H2, H4 AND H6. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 180390 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT 180390 AND DEVICE PART NUMBER 195-430H / LOT 173418. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. THE CURRENT OVERALL INCIDENT RATE FOR FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) FOR THIS SPECIFIC LOT BASED ON THE TOTAL QUANTITY OF DEVICES MANUFACTURED FOR DISTRIBUTION IS 0.001%. IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 SELF-TEST PERFORMED ON (B)(6) 2022 ON A NASAL SWAB. REPEAT TESTING WAS PERFORMED ON AN UNKNOWN DEVICE AND GENERATED A POSITIVE RESULT. ALTHOUGH REQUESTED, NO ADDITIONAL PATIENT INFORMATION INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1230258 BINAXNOW COVID-19 SELF-TEST LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 180390 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 Unknown