FDA Adverse Event Other Summary report: N

*

MDR report key: 1359225 · Received March 24, 2009

Report

Report Number
9610617-2009-00007
Event Type
Other
Date Received
March 24, 2009
Manufacturer
*
Product Code
GCJ
Report Source
Manufacturer report
Reporter Occupation
INVALID DATA

Narratives

Additional Manufacturer Narrative · 1

UNIT NOT RELEASED BY HOSP BUT WAS TESTED/INSPECTED BY THEIR BIOMED. HE TESTED UNIT WITH SETTINGS THEY KNEW THE OR USED AND THEY FOUND THAT THE THERMOFLATOR FUNCTIONED WITHIN SPECS; THERE WAS NO PROBLEM FOUND WITH THE UNIT. THE ALARM ACTIVATED WHEN IT WAS SUPPOSED TO. BIOMED STATED THAT THE SET PRESSURE WAS 15 AND THE FLOW AT 29. BIOMED STATED THAT IT WAS POSSIBLE THAT SOMEONE MADE AN ADJUSTMENT TO THE UNIT DURING PROCEDURE; HE ALSO BELIEVES THAT UNIT PROBABLY DID NOT ALARM WHEN IT HIT 30 BECAUSE IT DID NOT STAY THERE LONG ENOUGH (3-4 SECONDS) TO ACTIVATE THE ALARM. THE INSTRUCTION MANUAL SUPPORTS THE HOSPITAL'S CONTENTION: OVERPRESSURE ALARM - IF THE PT PRESSURE EXCEEDS THE SETPOINT BY 5MMHG FOR MORE THAN 5.0 SECONDS, AN AUDIBLE ALARM CONSISTING OF A SERIES OF 3 LONG BEEPS WILL SOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 * * GCJ * * *

Patients

Seq Age Sex Outcome Treatment
1 *