E360 VENTILATOR
Report
- Report Number
- 2023050-2009-00010
- Event Type
- Other
- Date Received
- April 7, 2009
- Date of Event
- March 5, 2009
- Report Date
- March 8, 2009
- Manufacturer
- NEWPORT MEDICAL INSTRUMENTS, INC.
- Product Code
- CBK
- PMA / PMN Number
- K053502
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVALUATION SUMMARY: ANALYSIS: TURNED THE VENTILATOR ON IN DIAGNOSTIC MODE AND OBSERVED THAT MACHINE 1 PRESSURE SHOWED 149.72 AND THE SEVEN SEGMENTS DISPLAYS WERE ALL OFF. REMOVED THE UNIT TOP COVER AND VISUALLY INSPECTED THE INNER CHAMBER FOR NOTICEABLE DEFECTS; NO PROBLEM WAS FOUND. USED AN OSCILLOSCOPE TO TAKE VOLTAGE MEASUREMENTS OF MACHINE 1 TRANSDUCER XD105 WITHOUT AIR PRESSURE APPLIED TO THE TRANSDUCER. MEASURED 3.6 VOLTS ON XD105 PIN 2 AND 2.68 VOLTS ON PIN 4 WHICH IS INCORRECT. BOTH PINS SHOULD HAVE MEASURED APPROXIMATELY 2.50 VOLTS. LIFTED PINS 2 AND 3 OF OPERATIONAL AMPLIFIER U126 TO ISOLATE XD105. MEASURED 3.50 VOLTS ON XD105 PIN 2 AND 2.53 VOLTS ON PIN 4 WHICH CONFIRMED THAT XD105 WAS DEFECTIVE. NO OTHER DEFECTS WERE FOUND ON THE ANALOG BOARD PCB2104A. CONCLUSION: PROBLEM CONFIRMED. MACHINE 1 TRANSDUCER XD105 LOCATED ON ANALOG BOARD PCB 2104A DOES NOT DELIVER THE CORRECT BIAS VOLTAGE OF 2.50 VOLTS ON PIN 2; IT DELIVERS 3.50 VOLTS. TRANSDUCER XD105 ON PCB2104A WAS FORWARDED TO MANUFACTURER FOR A DETAILED EVALUATION OF THE DIE. THE MANUFACTURER'S INVESTIGATION REVEALED THAT THE SENSOR EXHIBITED INTERMITTENT ELECTRICAL BEHAVIOR. THE INTERMITTENT BEHAVIOR OF THE SENSOR WAS VERY LIKELY TO BE CAUSED BY MICRO FRACTURES ON THE CERAMIC DURING THE PCB ASSEMBLY. THE HARD EPOXY CAN APPLY STRESS ON THE TRANSDUCER LEADS AND CERAMIC. THIS STRESS CAN CAUSE MICRO CRACKS AND AFFECT PERFORMANCE. AS A CORRECTION, THE SPECIFIED SILICON WAS INSTRUCTED TO USE TO SECURE THE TRANSDUCER PINS TO PCB2104A.
REPORTEDLY, THE VENTILATOR STOPPED VENTILATION DURING USE ON A PATIENT WITH DEVICE ALERT (AUDIBLE ALARM) AND ERROR MESSAGE "COMMUNICATION ERROR" ON THE DISPLAY. THE PATIENT'S SATURATION LEVEL WAS TEMPORARILY DROPPED. THE PATIENT WAS MANUALLY VENTILATED FOR 30 MINUTES AND THEN SWITCHED TO ANOTHER VENTILATOR. THE PATIENT'S SATURATION LEVEL WAS BACK TO NORMAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | E360 VENTILATOR | CBK | NEWPORT MEDICAL INSTRUMENTS, INC. | E360 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |