PENTAX
Report
- Report Number
- 9610877-2022-51099
- Event Type
- Malfunction
- Date Received
- February 24, 2022
- Date of Event
- February 7, 2022
- Report Date
- February 24, 2022
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FDT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PO
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL ED34-I10T-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K163614. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OBJECTIVE PRISM FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE OBJECTIVE PRISM. IN ADDITION, WE CONFIRMED THAT THE LCB DISTAL COVER GLASS BROKEN, THE AIR/WATER NOZZLE CLOGGED, THE SEGMENT LOOSENED, THE BENDING RUBBER WORN OUT, AND THE SEGMENT STEEL BRAID RUPTURE; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL ED34-I10T-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K163614. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OBJECTIVE PRISM FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE OBJECTIVE PRISM. IN ADDITION, WE CONFIRMED THAT THE LCB DISTAL COVER GLASS BROKEN, THE AIR/WATER NOZZLE CLOGGED, THE SEGMENT LOOSENED, THE BENDING RUBBER WORN OUT, AND THE SEGMENT STEEL BRAID RUPTURE; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 119968 | PENTAX | VIDEO DUODENOSCOPE - I10 STANDARD | FDT | HOYA CORPORATION PENTAX TOKYO OFFICE | ED34-I10T |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |