FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 13589691 · Received February 24, 2022

Report

Report Number
9610877-2022-51099
Event Type
Malfunction
Date Received
February 24, 2022
Date of Event
February 7, 2022
Report Date
February 24, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL ED34-I10T-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K163614. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OBJECTIVE PRISM FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE OBJECTIVE PRISM. IN ADDITION, WE CONFIRMED THAT THE LCB DISTAL COVER GLASS BROKEN, THE AIR/WATER NOZZLE CLOGGED, THE SEGMENT LOOSENED, THE BENDING RUBBER WORN OUT, AND THE SEGMENT STEEL BRAID RUPTURE; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL ED34-I10T-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K163614. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OBJECTIVE PRISM FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE OBJECTIVE PRISM. IN ADDITION, WE CONFIRMED THAT THE LCB DISTAL COVER GLASS BROKEN, THE AIR/WATER NOZZLE CLOGGED, THE SEGMENT LOOSENED, THE BENDING RUBBER WORN OUT, AND THE SEGMENT STEEL BRAID RUPTURE; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
119968 PENTAX VIDEO DUODENOSCOPE - I10 STANDARD FDT HOYA CORPORATION PENTAX TOKYO OFFICE ED34-I10T

Patients

Seq Age Sex Outcome Treatment
1 Unknown