FDA Adverse Event Injury Summary report: N

MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MMOL/L

MDR report key: 13589503 · Received February 24, 2022

Report

Report Number
2032227-2022-117639
Event Type
Injury
Date Received
February 24, 2022
Date of Event
February 13, 2022
Report Date
May 20, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000384272
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 0

(B)(4). MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. THE INFORMATION RELATED TO EVENT HAS BEEN UPDATED IN SUMMARY AND PROVIDED IN B5 SECTION OF THIS REPORT.

Additional Manufacturer Narrative · 0

RETAINER RING = BLACK. ON FEB 13, 2022, THE CUSTOMER ALLEGED THE INSULIN PUMP OVER DELIVERY AND HOSPITALIZED FOR LOW BG. THE TEST P-CAP LOCKS PROPERLY IN PLACE IN THE RESERVOIR COMPARTMENT NOTED. DEVICE RECEIVED WITH PILLOWING KEYPAD OVERLAY DURING THE VISUAL INSPECTION. THUMP AND CARELINK SOFTWARE WAS UTILIZED AND DOWNLOADED TRACE/HISTORY FILES PROPERLY. IN FURTHER FULL REVIEW OF THE DEVICE HISTORY ON THE EVENT DATE OF FEB 13, 2022, FOUND MULTIPLE BOLUS DELIVERIES THAT THE CUSTOMER MANUALLY PROGRAMMED AND THE DAILY TOTAL OF BOLUS INSULIN DELIVERED ARE 21.4 U. PUMP PASSED THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND DELIVERY ACCURACY TEST AT 0.08720 INCHES. DEVICE WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO MOISTURE OR COMPONENT DAMAGE ON THE ELECTRONICS, FORCE SENSOR AND MOTOR ASSEMBLY NOTED. THE FORCE SENSOR OFFSET MEASURED (23.1 MV). THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE ON THE NGP STB3 AND PASSED. IN SUMMARY, INSULIN PUMP PASSED ALL REQUIRED TESTING. UNABLE TO VERIFY CUSTOMER COMPLAINT FOR LOW BG. CUSTOMER ALLEGED FOR THE INSULIN PUMP OVER DELIVERY WAS NOT CONFIRMED. THE FORCE SENSOR IS WITHIN SPECIFICATION AND THE MOTOR FUNCTIONING PROPERLY. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS ALLEGED OVER DELIVERY. CUSTOMER REPORTED THAT THEY WERE HOSPITALIZED DUE TO LOW BLOOD GLUCOSE. THE BLOOD GLUCOSE VALUE WAS 1.8 MMOL/L. THE LOW BLOOD GLUCOSE WAS TREATED WITH GLUCAGON. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL CONTINUE USE OF THE DEVICE.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE AMBULANCE WAS CALLED AND THAT THEY WERE ADMITTED TO THE HOSPITAL ON (B)(6) 2022 DUE TO LOW BLOOD GLUCOSE. USER WAS HOSPITALIZED FOR 1 NIGHT. BLOOD GLUCOSE AT TIME OF HOSPITAL ADMISSION WAS 1.8 MMOL/L. BLOOD GLUCOSE AT TIME OF CALL WAS 7 MMOL/L. USER STATED THAT HER SPOUSE SAW HER FACE WAS WHITE AND SPOUSE COULD NOT WAKE HER UP. USER WAS TREATED WITH GLUCAGON AT HOSPITAL. PER CUSTOMER OVER DELIVERY WAS ALLEGED. USER STATED THEY HAD SENSOR UPDATING DURING THE NIGHT. DURING TROUBLESHOOTING CUSTOMER STATED THEY DO NOT DISCONNECT WHEN DOING A REWIND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1592172 MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MMOL/L ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1885 HG4Q92DZZ 000000763000384272

Patients

Seq Age Sex Outcome Treatment
1 Unknown