FDA Adverse Event Malfunction Summary report: N

BD SYRINGE LUER-LOK¿ 30 ML BULK PACK LUER LOCK TIP

MDR report key: 13586839 · Received February 23, 2022

Report

Report Number
1911916-2022-00097
Event Type
Malfunction
Date Received
February 23, 2022
Date of Event
December 10, 2021
Report Date
February 28, 2022
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY IT WAS REPORTED THERE WERE MULTIPLE DEFECTS WITH THE PRODUCTS RECEIVED. TO AID IN THE INVESTIGATION, FOURTEEN SAMPLES AND FIVE PHOTOS WERE RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. FIVE SAMPLES HAVE EMBEDDED DEGRADED RESIN. THE FIVE PHOTOS PROVIDED SHOW SOME OF THE SAMPLES RECEIVED. THE EMBEDDED DEGRADED RESIN IN THE COMPONENT TYPICALLY OCCURS AT THE STARTUP OR INTERMITTENTLY DURING THE INJECTION MOLDING PROCESS. THE DEGRADED RESIN CAN BREAK LOOSE AND BE MOLDED INTO COMPONENTS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 301033, LOT NUMBER 1180548. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD SYRINGE LUER-LOK¿ 30 ML BULK PACK LUER LOCK TIP APPROXIMATELY 241 PIECES HAD FOREIGN MATTER IN THE FLUID PATH. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 301033 BATCH NO: 1180548. IT WAS REPORTED BLACK DOTS. BLACK DOTS: 241 PIECES.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD SYRINGE LUER-LOK¿ 30 ML BULK PACK LUER LOCK TIP APPROXIMATELY 241 PIECES HAD FOREIGN MATTER IN THE FLUID PATH. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 301033 BATCH NO: 1180548 . IT WAS REPORTED BLACK DOTS BLACK DOTS: 241 PIECES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
848537 BD SYRINGE LUER-LOK¿ 30 ML BULK PACK LUER LOCK TIP PISTON SYRNIGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 1180548

Patients

Seq Age Sex Outcome Treatment
1 Unknown