FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 13575132 · Received February 22, 2022

Report

Report Number
9610877-2022-51044
Event Type
Malfunction
Date Received
February 22, 2022
Date of Event
February 14, 2022
Report Date
February 22, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL FNL-10RP3 IS AVAILABLE IN THE USA WITH A 510K NUMBER K951196. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CFB FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CFB. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1146771 PENTAX FIBER NASO PHARYNGO LARYNGOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE FNL-10RBS

Patients

Seq Age Sex Outcome Treatment
1 Unknown