DREAMSTATION AUTO CPAP
Report
- Report Number
- 2518422-2022-05349
- Event Type
- Malfunction
- Date Received
- February 21, 2022
- Date of Event
- August 4, 2021
- Report Date
- July 20, 2023
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- BZD
- PMA / PMN Number
- K131982
- Removal / Correction Number
- RES 88058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER PREVIOUSLY REPORTED AN ALLEGATION OF AN ISSUE RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THE MANUFACTURER RECEIVED INFORMATION ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE WAS RETURNED TO THE MANUFACTURER'S SERVICE CENTER FOR FURTHER EVALUATION. THE DEVICE WAS EVALUATED. THERE WAS NO MENTION OF VISUAL FINDINGS TO THE EXTERNAL PART OF THE DEVICE. THE INTERNAL ASPECT OF THE DEVICE WAS INSPECTED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE MANUFACTURER. THERE WERE NO ERRORS FOUND. THE MANUFACTURER CONCLUDES THAT THEY COULD NOT CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WAS NO VISIBLE FOAM DEGRADATION. UNIT SCRAPPED DUE TO AGE. SECTION H6 UPDATED IN THIS REPORT.
THE MANUFACTURER PREVIOUSLY SUBMITTED MDR 2518422-2022-05349-1 INCORRECTLY; FOLLOW-UP IS NOT REQUIRED FOR THIS MDR.
THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A DEVICE'S SOUND ABATEMENT FOAM. DURING DEVICE EVALUATION, WATER CONTAMINATION WAS OBSERVED IN THE DEVICE BLOWER RESULTING IN BLOWER BOX REPLACEMENT. PHYSICAL DAMAGE TO THE TOP ENCLOSURE NEAR AMBIENT LIGHT SENSOR. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1592527 | DREAMSTATION AUTO CPAP | VENTILATOR, NON-CONTINUOUS (RESPIRATOR) | BZD | RESPIRONICS, INC. | DSX500T11C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |