FDA Adverse Event Injury Summary report: N

BLT Ø4.1MM RC, SLA® 8MM, TIZR, LOXIM

MDR report key: 13570404 · Received February 21, 2022

Report

Report Number
0009613348-2022-16881
Event Type
Injury
Date Received
February 21, 2022
Date of Event
November 25, 2021
Report Date
February 21, 2022
Manufacturer
INSTITUT STRAUMANN AG
Product Code
DZE
UDI-DI
07630031707123
PMA / PMN Number
K150938
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE BATCH NUMBER COULD BE VERIFIED. OUR MANUFACTURING Q-SYSTEM ASSURES, THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFIRM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED.  THE NON-INTEGRATION OF AN ENDOSSEOUS DENTAL IMPLANT DURING THE HEALING PHASE IS A KNOWN INHERENT RISK OF THE TREATMENT WITH DENTAL IMPLANTS. MOST IMPLANT FAILURES OCCUR BEFORE OCCLUSAL LOADING (ANALAINEN ET AL. 2009). BASED ON CLINICAL STUDIES ABOUT 1-3% OF THE IMPLANTS FAIL WITHIN THE FIRST YEAR AFTER IMPLANTATION (GANELES ET AL. 2008). IMPLANTS MAY HAVE TO BE REMOVED IN CASE ONE OR MORE OF THE IMPLANT SUCCESS CRITERIA ARE NOT MET. IMPLANT SUCCESS CRITERIA ACCORDING TO BUSER ET AL. (1991) ARE: ABSENCE OF PERSISTENT SUBJECTIVE COMPLAINTS SUCH AS PAIN, FOREIGN BODY SENSATION AND /OR DYSESTHESIA. ABSENCE OF A RECURRENT PERI-IMPLANT INFECTION WITH SUPPURATION. ABSENCE OF IMPLANT MOBILITY. ABSENCE OF A CONTINUOUS RADIOLUCENCY AROUND THE IMPLANT. THE MANUFACTURER'S TREND ANALYSIS CONFIRMS THAT THE REPORTED EARLY FAILURE RATE ASSOCIATED WITH ITS DENTAL IMPLANTS IS BELOW THE EXPECTED FAILURE RATE FOR THIS TREATMENT AS PUBLISHED IN THE SCIENTIFIC LITERATURE.

Description of Event or Problem · 0

THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED 2021-11-12 IN FDI 36. ON 2021-11-25, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, BLEEDING AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1592572 BLT Ø4.1MM RC, SLA® 8MM, TIZR, LOXIM ENDOSSEOUS DENTAL IMPLANT DZE INSTITUT STRAUMANN AG SLA Roxolid BL AJV78 07630031707123

Patients

Seq Age Sex Outcome Treatment
1 62 YR Male Required Intervention