IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM
Report
- Report Number
- 0002023141-2022-00443
- Event Type
- Injury
- Date Received
- February 21, 2022
- Date of Event
- December 10, 2021
- Report Date
- July 19, 2022
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019522
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMMERBIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. PMA/510(K) NUMBER: K011028 AND K013227.
ZIMMERBIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. PMA/510(K) NUMBER: K011028 AND K013227.
ZIMMERBIOMET COMPLAINT NUMBER (B)(4). THE FOLLOWING SECTIONS HAVE BEEN UPDATED: D9: DEVICE AVAILABLE FOR EVALUATION CHANGE ¿NO' TO 'YES'. H3: DEVICE EVALUATED BY MANUFACTURER: CHANGE ¿NO' TO 'YES'. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1232211). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1232211) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED.
IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO BONE LOSS. EVENT WAS NOTICED AFTER 2 YEARS FROM THE IMPLANT PLACEMENT AND PROSTHETIC REHABILITATION.
IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO BONE LOSS. EVENT WAS NOTICED AFTER 2 YEARS FROM THE IMPLANT PLACEMENT AND PROSTHETIC REHABILITATION.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1456912 | IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSVB13 | 1232211 | 00889024019522 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Female | Required Intervention |