FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM

MDR report key: 13567097 · Received February 21, 2022

Report

Report Number
0002023141-2022-00443
Event Type
Injury
Date Received
February 21, 2022
Date of Event
December 10, 2021
Report Date
July 19, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019522
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMERBIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. PMA/510(K) NUMBER: K011028 AND K013227.

Additional Manufacturer Narrative · 0

ZIMMERBIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. PMA/510(K) NUMBER: K011028 AND K013227.

Additional Manufacturer Narrative · 0

ZIMMERBIOMET COMPLAINT NUMBER (B)(4). THE FOLLOWING SECTIONS HAVE BEEN UPDATED: D9: DEVICE AVAILABLE FOR EVALUATION CHANGE ¿NO' TO 'YES'. H3: DEVICE EVALUATED BY MANUFACTURER: CHANGE ¿NO' TO 'YES'. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1232211). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1232211) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO BONE LOSS. EVENT WAS NOTICED AFTER 2 YEARS FROM THE IMPLANT PLACEMENT AND PROSTHETIC REHABILITATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO BONE LOSS. EVENT WAS NOTICED AFTER 2 YEARS FROM THE IMPLANT PLACEMENT AND PROSTHETIC REHABILITATION.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1456912 IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 13MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVB13 1232211 00889024019522

Patients

Seq Age Sex Outcome Treatment
1 48 YR Female Required Intervention