FDA Adverse Event Malfunction Summary report: N

BD¿ PHLEBOTOMY SHARPS COLLECTOR, 1.5 QT NEEDLE PORT

MDR report key: 13562879 · Received February 18, 2022

Report

Report Number
2243072-2022-00249
Event Type
Malfunction
Date Received
February 18, 2022
Date of Event
February 9, 2022
Report Date
April 29, 2022
Manufacturer
BECTON DICKINSON
Product Code
FMI
UDI-DI
00382903054879
PMA / PMN Number
K943139
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE MANUFACTURING LOCATION FOR THIS PRODUCT IS (B)(4). THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD CORPORATE HEADQUARTERS IN (B)(4) HAS BEEN LISTED AND THE (B)(4) FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION: PICTURES WERE RECEIVED FROM THE CUSTOMER SHOWING THE DEFECT ON THE LIDS. THE DHR REVIEW WAS PERFORMED FOR LOT NUMBER PROVIDED AND A REVIEW OF THE HISTORICAL NCMR¿S FOR THE AFFECTED COMPONENT (LID) UNDER PART NUMBER BDX-8990020 WAS MADE AND IT WAS FOUND THAT 6 NCS WERE GENERATED FOR THE SAME DEFECT IN THE LAST 5 YEARS, THE VERY LAST EVENT WAS REPORTED BACK IN THE LAST YEAR. THERE IS AN OBSTRUCTION BY EXCESS OF PLASTIC MATERIAL ON LID¿S SKIRT, WHICH SEEMS TO BE GENERATED BY A BROKEN INSERT IN THE MOLD. HENCE, THE LID CAN¿T FIT THE BASE WHEN TRYING TO ASSEMBLY BY END USER.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD¿ PHLEBOTOMY SHARPS COLLECTOR, 1.5 QT NEEDLE PORT THE LID WAS DIFFICULT TO ATTACH PROPERLY. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THIS IS A REPORT ABOUT A LID ISSUE OF SHARPS COLLECTOR. ACCORDING TO THE CUSTOMER'S REPORT, THE HCP ATTEMPTED TO ASSEMBLE THE LID WITH THE BASE BUT THE LID DIDN'T FIT THE BASE PROPERLY.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD¿ PHLEBOTOMY SHARPS COLLECTOR, 1.5 QT NEEDLE PORT THE LID WAS DIFFICULT TO ATTACH PROPERLY. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THIS IS A REPORT ABOUT A LID ISSUE OF SHARPS COLLECTOR. ACCORDING TO THE CUSTOMER'S REPORT, THE HCP ATTEMPTED TO ASSEMBLE THE LID WITH THE BASE BUT THE LID DIDN'T FIT THE BASE PROPERLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
14708 BD¿ PHLEBOTOMY SHARPS COLLECTOR, 1.5 QT NEEDLE PORT HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON 305487 1175904 00382903054879

Patients

Seq Age Sex Outcome Treatment
1 Unknown