FDA Adverse Event Malfunction Summary report: N

SOCKET WRENCH FOR VEPTR NUT

MDR report key: 13560766 · Received February 18, 2022

Report

Report Number
2939274-2022-00546
Event Type
Malfunction
Date Received
February 18, 2022
Date of Event
January 20, 2022
Report Date
February 18, 2022
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HXC
UDI-DI
10705034720128
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). ADDITIONAL NARRATIVE: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10: ADDITIONAL NARRATIVE: D4, E1, E3: INITIAL REPORTER IS A SYNTHES EMPLOYEE H3, H4, H6: PART 03.641.004, LOT J001347-09: RELEASE TO WAREHOUSE DATE: DECEMBER 04, 2020. SUPPLIER: AVALIGN TECHNOLOGIES-NEMCOMED. NO NON-CONFORMANCE REPORTS (NCRS) WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. H3, H6: A SERVICE AND REPAIR EVALUATION WAS COMPLETED: DURING EVALUATION AT SERVICE AND REPAIR, THE DEVICE WAS FOUND TO HAVE FAILED CALIBRATION. THE REPAIR TECHNICIAN REPORTED THE DEVICE FAILED HIGH. THE CAUSE OF THE ISSUE IS UNKNOWN. THE ITEM WILL BE REPAIRED AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. THE EVALUATION WAS CONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H11: CORRECTED DATA: D9.

Description of Event or Problem · 0

IT WAS REPORTED ON (B)(6), 2022, DURING EVALUATION AT SERVICE AND REPAIR, THE SOCKET WRENCH FOR VEPTR NUT WAS FOUND TO HAVE FAILED CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE SOCKET WRENCH FOR VEPTR NUT. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
6445 SOCKET WRENCH FOR VEPTR NUT WRENCH HXC WRIGHTS LANE SYNTHES USA PRODUCTS LLC 03.641.004 J001347-09 10705034720128

Patients

Seq Age Sex Outcome Treatment
1 Unknown