SOCKET WRENCH FOR VEPTR NUT
Report
- Report Number
- 2939274-2022-00546
- Event Type
- Malfunction
- Date Received
- February 18, 2022
- Date of Event
- January 20, 2022
- Report Date
- February 18, 2022
- Manufacturer
- WRIGHTS LANE SYNTHES USA PRODUCTS LLC
- Product Code
- HXC
- UDI-DI
- 10705034720128
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
PRODUCT COMPLAINT # (B)(4). ADDITIONAL NARRATIVE: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10: ADDITIONAL NARRATIVE: D4, E1, E3: INITIAL REPORTER IS A SYNTHES EMPLOYEE H3, H4, H6: PART 03.641.004, LOT J001347-09: RELEASE TO WAREHOUSE DATE: DECEMBER 04, 2020. SUPPLIER: AVALIGN TECHNOLOGIES-NEMCOMED. NO NON-CONFORMANCE REPORTS (NCRS) WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. H3, H6: A SERVICE AND REPAIR EVALUATION WAS COMPLETED: DURING EVALUATION AT SERVICE AND REPAIR, THE DEVICE WAS FOUND TO HAVE FAILED CALIBRATION. THE REPAIR TECHNICIAN REPORTED THE DEVICE FAILED HIGH. THE CAUSE OF THE ISSUE IS UNKNOWN. THE ITEM WILL BE REPAIRED AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. THE EVALUATION WAS CONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H11: CORRECTED DATA: D9.
IT WAS REPORTED ON (B)(6), 2022, DURING EVALUATION AT SERVICE AND REPAIR, THE SOCKET WRENCH FOR VEPTR NUT WAS FOUND TO HAVE FAILED CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE SOCKET WRENCH FOR VEPTR NUT. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 6445 | SOCKET WRENCH FOR VEPTR NUT | WRENCH | HXC | WRIGHTS LANE SYNTHES USA PRODUCTS LLC | 03.641.004 | J001347-09 | 10705034720128 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |