CUP: MPACT ACETABULAR SHELL Ø50 TWO-HOLES
Report
- Report Number
- 3005180920-2022-00137
- Event Type
- Injury
- Date Received
- February 18, 2022
- Date of Event
- January 27, 2022
- Report Date
- February 18, 2022
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LPH
- UDI-DI
- 07630030810763
- PMA / PMN Number
- K132879
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 02 FEBRUARY 2022: LOT 188867: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 27-FEB-2019. EXPIRATION DATE: 2024-02-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ANOTHER SIMILAR REPORTED EVENT. ADDITIONAL IMPLANT INVOLVED: LINER: MPACT 01.32.3241HCT FLAT PE HC LINER Ø32/D (K103721) LOT. 2000178. BATCH REVIEW PERFORMED ON 02 FEBRUARY 2022: LOT 2000178: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06-MAR-2020. EXPIRATION DATE: 2025-02-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED. ADDITIONAL IMPLANT INVOLVED: STEM: AMISTEM P 01.18.402 AMISTEM-P STD STEM SIZE 2 (K173794) LOT. 2000050. BATCH REVIEW PERFORMED ON 02 FEBRUARY 2022: LOT 2000050: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-MAY-2020. EXPIRATION DATE: 2025-04-29. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED. ADDITIONAL IMPLANT INVOLVED: BALL HEADS: MECTACER 01.29.205 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 32 SIZE M 0 (K112115) LOT. 1904642. BATCH REVIEW PERFORMED ON 02 FEBRUARY 2022: LOT 1904642: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 19-SEP-2019. EXPIRATION DATE: 2024-09-09. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL IMPLANT INVOLVED: SCREWS: MPACT 01.32.6530 CANCELLOUS BONE SCREW, FLAT HEAD Ø 6,5 L 30 (K103721) LOT. 2002265. BATCH REVIEW PERFORMED ON 02 FEBRUARY 2022: LOT 2002265: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 30-APR-2020. EXPIRATION DATE: 2025-04-20. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH OTHER 2 SIMILAR REPORTED EVENTS.
THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. AT 9 MONTHS AFTER THE PRIMARY SURGERY THE SURGEON PERFORMED A WASHOUT, REMOVED ALL IMPLANTS, AND IMPLANTED AN ANTIBIOTIC SPACER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 473014 | CUP: MPACT ACETABULAR SHELL Ø50 TWO-HOLES | CEMENTLESS ACETABULAR CUP | LPH | MEDACTA INTERNATIONAL SA | 01.32.150DH | 188867 | 07630030810763 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | Required Intervention |