FDA Adverse Event Malfunction Summary report: N

BD VERITOR ¿ AT-HOME COVID-19 TEST

MDR report key: 13552850 · Received February 17, 2022

Report

Report Number
1119779-2022-00257
Event Type
Malfunction
Date Received
February 17, 2022
Date of Event
January 24, 2022
Report Date
July 1, 2022
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
QKP
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

EUA# (B)(4). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: THIS SUMMARIZES THE INVESTIGATION RESULTS REGARDING A COMPLAINT THAT ALLEGES THE BD VERITOR AT HOME COVID-19 TEST (MATERIAL # 256094), BATCH NUMBER 2006644, ¿HAD A NEGATIVE RESULT HOWEVER PCR RESULT WAS POSITIVE.¿ BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE ALL COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. AN INVESTIGATION AND TESTING WERE PERFORMED ON THE BATCH NUMBER PROVIDED. RESULTS WERE ACCEPTABLE AND NO RELEVANT ISSUE WAS FOUND. THE COMPLAINT WAS UNABLE TO BE CONFIRMED. THE ROOT CAUSE COULD NOT BE IDENTIFIED. CURRENTLY, THERE ARE NO ADVERSE TRENDS IDENTIFIED FOR FALSE NEGATIVE. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE TESTING FOR SARS COV-2 WITH BD VERITOR ¿ AT-HOME COVID-19 TEST A FALSE NEGATIVE RESULT WAS OBTAINED. CONFIRMATORY TESTING WAS PERFORMED USING PCR TEST METHOD AND THE RESULT WAS POSITIVE. THERE WAS NO REPORT OF USER IMPACT. EUA# (B)(4). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER TOOK A PCR TEST AND TESTED POSITIVE BUT TESTED NEGATIVE WITH OUR TEST.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE TESTING FOR SARS COV-2 WITH BD VERITOR ¿ AT-HOME COVID-19 TEST A FALSE NEGATIVE RESULT WAS OBTAINED. CONFIRMATORY TESTING WAS PERFORMED USING PCR TEST METHOD AND THE RESULT WAS POSITIVE. THERE WAS NO REPORT OF USER IMPACT. EUA# 210417. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER TOOK A PCR TEST AND TESTED POSITIVE BUT TESTED NEGATIVE WITH OUR TEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1298756 BD VERITOR ¿ AT-HOME COVID-19 TEST NOT CLASSIFIED QKP BECTON, DICKINSON & CO. (SPARKS) 2006644

Patients

Seq Age Sex Outcome Treatment
1 Unknown