CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Report
- Report Number
- 2249723-2022-00316
- Event Type
- Malfunction
- Date Received
- February 17, 2022
- Date of Event
- January 21, 2022
- Report Date
- July 31, 2023
- Manufacturer
- DATASCOPE CORP. - MAHWAH
- Product Code
- DSP
- UDI-DI
- 10607567112541
- PMA / PMN Number
- K063525
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
TESTING OF ACTUAL/SUSPECTED DEVICE/10: A GETINGE FIELD SERVICE ENGINEER (FSE) EVALUATED THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT AND WAS ABLE TO REPRODUCE THE REPORTED ISSUE. THE FSE DETERMINED THAT THE POWER SUPPLY WAS DEFECTIVE. TO FIX THE ISSUE, THE FSE REPLACED THE POWER SUPPLY, AND PERFORMED ALL FUNCTIONAL AND SAFETY CHECKS TO MEET FACTORY SPECIFICATIONS. THE IABP WAS THEN RELEASED TO THE CUSTOMER AND CLEARED FOR CLINICAL SERVICE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.
CORRECTED FIELDS: D1, D4, H6 (TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS) A GETINGE FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE IABP UNIT AND WAS ABLE TO REPRODUCE THE REPORTED ISSUE. THE FSE REPORTED THAT THE POWER SUPPLY WAS DEFECTIVE SO THE FSE REPLACED THE POWER SUPPLY FROM COMPANY STOCK TO HOSPITAL UNDER CMC AND THE MACHINE WAS WORKING FINE. THE FSE COMPLETED ALL SAFETY, FUNCTIONALITY, AND CALIBRATION CHECKS AND ALL TESTS PASSED TO FACTORY SPECIFICATIONS. THE IABP WAS THEN RELEASED TO THE CUSTOMER AND CLEARED FOR CLINICAL SERVICE.
IT WAS REPORTED BY THE CUSTOMER THAT DURING A ROUTINE CHECK THE CS300 INTRA- AORTIC BALLOON PUMP (IABP) UNIT WILL NOT POWER ON. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.
N/A.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 999016 | CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. - MAHWAH | 0998-00-3023-55 | N/A | 10607567112541 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |