IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 8MM
Report
- Report Number
- 0002023141-2022-00383
- Event Type
- Injury
- Date Received
- February 17, 2022
- Date of Event
- November 14, 2021
- Report Date
- July 19, 2022
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. ADDITIONAL DEVICE INFORMATION UNKNOWN / NOT PROVIDED. EMAIL ADDRESS UNKNOWN / NOT PROVIDED. PREMARKET IDENTIFICATION K011028 AND K013227. DEVICE MANUFACTURER DATE UNKNOWN / NOT PROVIDED.
ZIMMERBIOMET COMPLAINT NUMBER (B)(4). THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT. B5: DESCRIBE EVENT OR PROBLEM. D4: EXPIRATION DATE . D9: DEVICE AVAILABLE FOR EVALUATION CHANGE ¿NO' TO 'YES'. G3: DATE RECEIVED BY THE MANUFACTURER. H1: TYPE OF REPORT, FOLLOW-UP NUMBER. H2: FOLLOW UP TYPE. H3: DEVICE EVALUATED BY MANUFACTURER: CHANGE ¿NO' TO 'YES'. H4: DEVICE MANUFACTURE. H6: EVALUATION CODES. H10: ADDITIONAL NARRATIVE. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1230741). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1230741) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED.
ZIMMERBIOMET COMPLAINT NUMBER (B)(4). THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT. B5: DESCRIBE EVENT OR PROBLEM. D4: LOT NUMBER . D4: EXPIRATION DATE AND UDI NUMBER. G3: DATE RECEIVED BY THE MANUFACTURER. H1: TYPE OF REPORT, FOLLOW-UP NUMBER. H2: FOLLOW UP TYPE. H4: DEVICE MANUFACTURE DATE. H10: ADDITIONAL NARRATIVE. SUPPLEMENTAL REPORT SUBMITTED AS LOT NUMBER HAS BEEN PROVIDED AFTER FOLLOW UP WITH THE DOCTOR.
IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION 26 (FDI) WAS REMOVED DUE TO BONE LOSS AFTER 2 YEARS. THE CROWN STARTED TO OSCILLATE.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1759912 | IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 8MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1230741 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Male | Required Intervention |