FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSION PUMP

MDR report key: 1354165 · Received March 31, 2009

Report

Report Number
6000001-2009-00288
Event Type
Malfunction
Date Received
March 31, 2009
Date of Event
March 1, 2009
Report Date
March 5, 2009
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: THE REPORTED CONDITION OF "PHM KEYPAD INOPERABLE" WAS CONFIRMED AS THE OPEN AND STOP KEYS WERE NOT FUNCTIONING. INSPECTION OF THE PUMP REVEALED THE OPEN AND STOP KEYS THAT WERE NOT FUNCTIONING WAS CAUSED BY A FAULTY PUMP HEAD MODULE (PHM) KEYPAD. THE PHM KEYPAD WAS REPLACED ON THE PUMP TO CORRECT THIS CONDITION. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE FACILITY REPORTED A PUMP WITH A "PHM KEYPAD INOPERABLE". THE CONDITION WAS CONFIRMED AS THE OPEN AND STOP KEYS WERE NOT FUNCTIONING. THE CONDITION OCCURRED DURING BIO-MED TESTING. THERE WAS NO PT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFO IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSION PUMP 80FRN FRN BAXTER HEALTHCARE PTE. LTD.

Patients

Seq Age Sex Outcome Treatment
1