FDA Adverse Event Malfunction Summary report: N

INFUSION PUMP

MDR report key: 1354159 · Received March 31, 2009

Report

Report Number
6000001-2009-00278
Event Type
Malfunction
Date Received
March 31, 2009
Date of Event
March 9, 2009
Report Date
March 12, 2009
Manufacturer
BAXTER HEALTHCARE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BAXTER HAS NOT RECEIVED THE DEVICE FOR EVALUATION. THE CUSTOMER RETURNED THE DEVICE TO THEIR FACILITY BIO-MED DEPARTMENT FOR TESTING IMMEDIATELY FOLLOWING THE REPORTED INCIDENT AND DETERMINED THAT THE DEVICE FUNCTIONED AS INTENDED. THE CUSTOMER STATED THE DEVICE WILL NOT BE RETURNED TO BAXTER FOR EVALUATION. A FOLLOW-UP REPORT WILL BE FILED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. THIS REPORT REPRESENTS ALL INFORMATION KNOWN BY THE REPORTER AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE NURSE SPIKED THE MINI-BAG AND PROGRAMMED THE FLO-GARD INFUSION PUMP, SERIAL NUMBER UNKNOWN, TO HAVE THE BAG RUN DRY. THEY DO THIS BY PROGRAMMING THE VTBI (VOLUME TO BE INFUSED) TO BE GREATER THAN THE AMOUNT OF SOLUTION IN THE BAG. WHEN THE PUMP COMPLETED THE INFUSION AND RAN THE BAG DRY, THE NURSE EITHER DISCOVERED OR WAS ALERTED BY THE PATIENT'S FAMILY MEMBER TO THE EMPTY BAG. THE BAG AND UPSTREAM TUBING WERE FOUND EMPTY. FLUID IS STILL PRESENT IN THE DOWNSTREAM TUBING AND APPEARS TO BE 'PULSATING". BLOOD WAS ALSO PULSATING AT THE PATIENT IV SITE. THE PUMP DID NOT ALARM AT THIS POINT. THE NURSE REMOVED THE TUBING FROM THE PUMP AND DISCOVERED THE TUBING IS EMPTY AND FLAT AND DOES NOT RETRACT AS EXPECTED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THIS CONDITION OCCURRED WHILE ADMINISTERING DECADRON TO A PATIENT FROM THE 50ML MINI-BAG VIAFLEX, PRODUCT LOT NUMBER P222414, IN 2009. THE CUSTOMER BELIEVES THIS ISSUE IS RELATED TO USE OF THIS 50ML MINI-BAG SINCE THEY BEGAN EXPERIENCING THIS ISSUE UPON INTRODUCING THIS PRODUCT INTO THEIR PROCESS. THIS CONDITION ONLY OCCURS WITH MINI-BAGS THAT ARE NOT PRE-PRIMED BY THE PHARMACY. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSION PUMP 80FRN FRN BAXTER HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1