FDA Adverse Event Malfunction Summary report: N

LIBRE 3 SENSOR FREESTYLE

MDR report key: 13541439 · Received February 16, 2022

Report

Report Number
2954323-2022-05311
Event Type
Malfunction
Date Received
February 16, 2022
Date of Event
February 1, 2022
Report Date
July 8, 2022
Manufacturer
ABBOTT DIABETES CARE LTD
Product Code
QLG
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

REPEATED ATTEMPTS BY ADC TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL AND/OR THE CUSTOMER DISCARDED THE PRODUCT. AT THIS TIME PRODUCT HAS NOT YET BEEN RETURNED. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. DHRS FOR THE FS LIBRE SENSOR AND FS LIBRE SENSOR KIT WERE REVIEWED AND THE DHRS SHOWED THE FS LIBRE SENSOR AND SENSOR KIT PASSED ALL TESTS PRIOR TO RELEASE. IF THE PRODUCT IS RETURNED, A PHYSICAL INVESTIGATION WILL BE PERFORMED AND A FOLLOW-UP REPORT SUBMITTED. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, MODEL NUMBER 72114-01 WHICH HAS A SIMILAR PRODUCT DISTRIBUTED IN THE U.S., MODEL NUMBER 71992-01. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

REPEATED ATTEMPTS BY ADC TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL AND/OR THE CUSTOMER DISCARDED THE PRODUCT. AT THIS TIME PRODUCT HAS NOT YET BEEN RETURNED. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. DHRS FOR THE FS LIBRE SENSOR AND FS LIBRE SENSOR KIT WERE REVIEWED AND THE DHRS SHOWED THE FS LIBRE SENSOR AND SENSOR KIT PASSED ALL TESTS PRIOR TO RELEASE. IF THE PRODUCT IS RETURNED, A PHYSICAL INVESTIGATION WILL BE PERFORMED AND A FOLLOW-UP REPORT SUBMITTED. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, MODEL NUMBER 72114-01 WHICH HAS A SIMILAR PRODUCT DISTRIBUTED IN THE U.S., MODEL NUMBER 71992-01. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

SENSOR (B)(6) HAS BEEN RETURNED AND INVESTIGATED. VISUAL INSPECTION WAS PERFORMED ON THE RETURNED SENSOR, NO ISSUES WERE OBSERVED. DATA WAS SUCCESSFULLY EXTRACTED FROM THE RETURNED SENSOR USING APPROVE SOFTWARE. SENSOR LOW SIGNAL WAS OBSERVED. THE SENSOR WAS SENT FOR FURTHER INVESTIGATION AND DE-CASES. VISUAL INSPECTION WAS PREFORMED AND NO ISSUE WERE OBSERVED. THIS ISSUE IS NOT CONFIRMED TO EARLY SENSOR REMOVAL. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
536812 LIBRE 3 SENSOR FREESTYLE CONTINUOUS GLUCOSE MONITORING SYSTEM QLG ABBOTT DIABETES CARE LTD 72114-01

Patients

Seq Age Sex Outcome Treatment
1 Unknown