SIGMA PLI GVF INS 4 10MM
Report
- Report Number
- 1818910-2022-02901
- Event Type
- Malfunction
- Date Received
- February 15, 2022
- Date of Event
- February 3, 2022
- Report Date
- February 15, 2022
- Manufacturer
- DEPUY ORTHOPAEDICS INC US
- Product Code
- NJL
- UDI-DI
- 10603295062615
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY : DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RECEIVED FOR EXAMINATION. ALL AVAILABLE PHOTOGRAPH EVIDENCE WAS REVIEWED AND FOUND THE PACKAGE HAD A CUT. IT CAN BE DETERMINED THAT DEVICE STERILITY WAS COMPROMISED DUE TO THE DAMAGE IDENTIFIED. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED, AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT : PRODUCT CODE 158104010, LOT NUMBER D21061369 WAS MANUFACTURED ON 21-JUN-2021. 44 PCS PARTS WERE MANUFACTURED PER SPECIFICATION AND ALL RAW MATERIAL MET SPECIFICATION. NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED. DEVICE HISTORY REVIEW : A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE(PRODUCT CODE: 158104010/LOT NUMBER: D21061369), AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED.
WHILST OPENING THE INSERT, THE SCRUB NURSE NOTICED A HOLE ABOUT 1CM LONG IN THE INNER PACKAGING OF THE IMPLANT. THE BOX AND CELLOPHANE SHOWED NO SIGNS OF DAMAGE. I HAVE ATTACHED PHOTOS OF THE 2 INNER PACKAGING. SURGICAL DELAY OF 5 MINUTES OCCURRED. NO ADVERSE AFFECTS ON PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 468171 | SIGMA PLI GVF INS 4 10MM | SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT | NJL | DEPUY ORTHOPAEDICS INC US | 1581-04-010 | D21061369 | 10603295062615 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |