I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2022-00027
- Event Type
- Malfunction
- Date Received
- February 15, 2022
- Date of Event
- February 9, 2022
- Report Date
- May 2, 2022
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- UDI-DI
- 10054749000163
- PMA / PMN Number
- K053110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT: # (B)(4). TYPOGRAPHIC ERRORS NOTED DURING A REVIEW ON 28-APR-2022. CORRECTED INFORMATION: B5 PRODUCT INFORMATION: CHANGE FROM CG8+ TO CHEM8+. D1 BRAND NAME: CHANGE FROM CG8+ TO CHEM8+. D2 COMMON DEVICE NAME: CHANGE FROM CG8+, CHL TO CHEM8+ JGS. G4 PREMARKET IDENTIFICATION: CHANGE FROM K940918 TO K053110.
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 23-FEB-2021. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE LOT PASSED FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AG, APPENDIX 1- PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. A PREVIOUS DEFICIENCY WAS IDENTIFIED FOR SUPPRESSED RESULTS AND IS BEING INVESTIGATED IN QUALITY RECORD (QR) (B)(4).
ON (B)(6) 2022, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CG8+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT HEMATOCRIT RESULT OF 29% PCV ON A PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION. I-STAT HCT 29% PCV, I-STAT CREA 0.7MG/DL, MANUAL HCT 45% PCV. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
NA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379611 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | H21357 | 10054749000163 |
| 669883 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | H21357 | 10054749000163 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |