FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 13536702 · Received February 15, 2022

Report

Report Number
2245578-2022-00027
Event Type
Malfunction
Date Received
February 15, 2022
Date of Event
February 9, 2022
Report Date
May 2, 2022
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
UDI-DI
10054749000163
PMA / PMN Number
K053110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). TYPOGRAPHIC ERRORS NOTED DURING A REVIEW ON 28-APR-2022. CORRECTED INFORMATION: B5 PRODUCT INFORMATION: CHANGE FROM CG8+ TO CHEM8+. D1 BRAND NAME: CHANGE FROM CG8+ TO CHEM8+. D2 COMMON DEVICE NAME: CHANGE FROM CG8+, CHL TO CHEM8+ JGS. G4 PREMARKET IDENTIFICATION: CHANGE FROM K940918 TO K053110.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 23-FEB-2021. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE LOT PASSED FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AG, APPENDIX 1- PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. A PREVIOUS DEFICIENCY WAS IDENTIFIED FOR SUPPRESSED RESULTS AND IS BEING INVESTIGATED IN QUALITY RECORD (QR) (B)(4).

Description of Event or Problem · 0

ON (B)(6) 2022, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CG8+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT HEMATOCRIT RESULT OF 29% PCV ON A PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION. I-STAT HCT 29% PCV, I-STAT CREA 0.7MG/DL, MANUAL HCT 45% PCV. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

Description of Event or Problem · 0

NA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379611 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JGS ABBOTT POINT OF CARE NA H21357 10054749000163
669883 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JGS ABBOTT POINT OF CARE NA H21357 10054749000163

Patients

Seq Age Sex Outcome Treatment
1 Unknown