FDA Adverse Event Malfunction Summary report: N

NEO METER OPTIUM FREESTYLE

MDR report key: 13534074 · Received February 15, 2022

Report

Report Number
2954323-2022-05090
Event Type
Malfunction
Date Received
February 15, 2022
Date of Event
February 11, 2022
Report Date
August 1, 2022
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
NBW
PMA / PMN Number
K140371
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CN
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

AN EXTENDED INVESTIGATION HAS BEEN CONDUCTED, WHICH INVOLVED A MANUFACTURING REVIEW (INCLUDING 5 YEARS OF DEVICE HISTORY RECORDS (DHRS)), PREVIOUS COMPLAINT AND CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATIONS CONDUCTED FOR BLANK SCREEN ISSUES, PROCESS FAILURE MODE EFFECTS ANALYSES (PFMEAS), DESIGN CONTROLS, AND DESIGN FAILURE MODE EFFECTS ANALYSES (DFMEAS), RISK MANAGEMENT REPORTS, RISK EVALUATIONS, AND LABEL COPY. THE INVESTIGATION DID NOT IDENTIFY ANY INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. IF THE PRODUCT IS RETURNED, AN INVESTIGATION WILL BE PERFORMED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

THE RETURNED METER WAS INVESTIGATED. VISUAL INSPECTION WAS PERFORMED ON THE RETURNED METER AND NO ISSUES WERE OBSERVED. INSERTED BATTERIES BUT INDICATOR LIGHTS DID NOT FLASH. METER POWERED ON WITH BUTTON DEPRESSION. BLACK SPOTS WAS OBSERVED. DSPLGHT-2 WAS CLONED TO ADDRESS THE ISSUE. BLANK SCREEN WAS NOT OBSERVED. THE COMPLAINT WAS NOT CONFIRMED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED

Additional Manufacturer Narrative · 0

AN EXTENDED INVESTIGATION HAS BEEN CONDUCTED, WHICH INVOLVED A MANUFACTURING REVIEW (INCLUDING 5 YEARS OF DEVICE HISTORY RECORDS (DHRS)), PREVIOUS COMPLAINT AND CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATIONS CONDUCTED FOR BLANK SCREEN ISSUES, PROCESS FAILURE MODE EFFECTS ANALYSES (PFMEAS), DESIGN CONTROLS, AND DESIGN FAILURE MODE EFFECTS ANALYSES (DFMEAS), RISK MANAGEMENT REPORTS, RISK EVALUATIONS, AND LABEL COPY. THE INVESTIGATION DID NOT IDENTIFY ANY INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. IF THE PRODUCT IS RETURNED, AN INVESTIGATION WILL BE PERFORMED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 0

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 0

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 0

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
875697 NEO METER OPTIUM FREESTYLE BLOOD GLUCOSE MONITORING SYSTEM NBW ABBOTT DIABETES CARE INC 75175-84

Patients

Seq Age Sex Outcome Treatment
1 Unknown