FDA Adverse Event Malfunction Summary report: N

MEGABITER, UP CURVED TIP, 5.5MM X 2.5MM

MDR report key: 13528661 · Received February 14, 2022

Report

Report Number
1220246-2022-04468
Event Type
Malfunction
Date Received
February 14, 2022
Date of Event
December 19, 2021
Report Date
April 13, 2022
Manufacturer
ARTHREX, INC.
Product Code
MDM
UDI-DI
00888867032743
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LH
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE DETERMINED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION. IF THE DEVICE BECOMES AVAILABLE FOR EVALUATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

COMPLAINT CONFIRMED. ONE UNPACKAGED AR-41026 WAS RECEIVED FOR INVESTIGATION. VISUAL INSPECTION IDENTIFIED THAT THE 723-4002 CUTTER HAD BROKEN OFF. WEAR WAS NOTED TO THE CUTTER FRAGMENT. A PROBABLE CAUSE IS ATTRIBUTED TO WEAR AND TEAR DAMAGE BASED ON THE AGE OF THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A SURGERY A PART OF THE DEVICE BROKE INSIDE THE PATIENT. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY WAS REPORTED. NO FURTHER INFORMATION RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
784131 MEGABITER, UP CURVED TIP, 5.5MM X 2.5MM INSTRUMENT, MANUAL, SURGICAL, GENERAL USE MDM ARTHREX, INC. MEGABITER, UP CURVED TIP, 5.5MM X 2.5MM 87030 00888867032743

Patients

Seq Age Sex Outcome Treatment
1 Unknown