FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 13524398 · Received February 14, 2022

Report

Report Number
3008642652-2022-01605
Event Type
Malfunction
Date Received
February 14, 2022
Date of Event
January 14, 2022
Report Date
August 16, 2022
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
UDI-DI
00855778005005
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION OF MONITOR IS UNDERWAY. THE REPORTED PROBLEM (SERVICE CODE 102) HAS BEEN CONFIRMED. UPON INVESTIGATION THE MONITOR WAS DISPLAYING A SERVICE CODE 102 AND FAILED INCOMING FUNCTIONAL TESTING. ROOT CAUSE INVESTIGATION IS UNDERWAY. NO ADVERSE EVENT RESULTED FROM THE DAMAGED MONITOR.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION OF MONITOR HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 102) HAS BEEN CONFIRMED. UPON INVESTIGATION THE MONITOR FAILED A PULSE TEST AND DISPLAYED A SERVICE CODE 102 (CHARGE PROFILE FAULT). THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN R45 RESISTOR ON THE COMPUTER/ANALOG BOARD. THE ROOT CAUSE FOR THE OPEN R45 RESISTOR COULD NOT BE POSITIVELY IDENTIFIED. THERE WAS NO ADVERSE EVENT THAT RESULTED FROM THE DSAMAGED MONITOR. DEVICE EVALUATION OF MONITOR IS UNDERWAY. THE REPORTED PROBLEM (SERVICE CODE 102) HAS BEEN CONFIRMED. UPON INVESTIGATION THE MONITOR WAS DISPLAYING A SERVICE CODE 102 AND FAILED INCOMING FUNCTIONAL TESTING. ROOT CAUSE INVESTIGATION IS UNDERWAY. NO ADVERSE EVENT RESULTED FROM THE DAMAGED MONITOR.

Description of Event or Problem · 0

A US DISTRIBUTOR RETURNED A MONITOR AND REPORTED THE MONITOR DISPLAYED A SERVICE CODE 102 (CHARGE PROFILE FAULT) AND FAILED INCOMING TESTING.

Description of Event or Problem · 0

A US DISTRIBUTOR RETURNED A MONITOR AND REPORTED THE MONITOR DISPLAYED A SERVICE CODE 102 (CHARGE PROFILE FAULT) AND FAILED INCOMING TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
784119 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 00855778005005

Patients

Seq Age Sex Outcome Treatment
1 Unknown