FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 13523772
·
Received February 14, 2022
Report
- Report Number
- 3013756811-2022-13199
- Event Type
- Malfunction
- Date Received
- February 14, 2022
- Date of Event
- January 30, 2022
- Report Date
- February 14, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
DEVICE NOT RETURNED.
Description of Event or Problem · 0
IT WAS REPORTED THAT AIR BUBBLES WERE OBSERVED IN THE INFUSION SET TUBING. REPORTEDLY, THE CUSTOMER LOADED CARTRIDGES WITH COLD INSULIN. TANDEM TECHNICAL SUPPORT (CTS) EDUCATED THE CUSTOMER REGARDING CARTRIDGE/INSULIN LABELING. CTS GUIDED THE CUSTOMER THROUGH PRIMING THE AIR BUBBLES OUT. CUSTOMER'S BLOOD GLUCOSE RANGED FROM 232-233 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 110664 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Male | INFUSION SET: VARISOFT, INSULIN: NOVOLOG |