FDA Adverse Event Injury Summary report: N

HAHN TAPERED IMPLANT 03.5 X11.5MM

MDR report key: 13522652 · Received February 14, 2022

Report

Report Number
3011649314-2022-00021
Event Type
Injury
Date Received
February 14, 2022
Date of Event
December 10, 2021
Report Date
June 3, 2024
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN RETURNED. IF/WHEN THERE IS MORE INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN RETURNED. HOWEVER, THE NON-VISUAL DEVICE EVALUATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR LOT# 6051418 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: THE PACKAGED STOCK PRODUCT IS NOT APPLICABLE FOR REVIEW SINCE NO DEFECT COULD BE CONFIRMED. INVESTIGATION METHODS/RESULTS: CUSTOMER HAS NOT RETURNED THE REPORTED DEVICE FOR REVIEW TO DATE. HOWEVER, THE NON-VISUAL DEVICE INVESTIGATION HAS BEEN COMPLETED. ROOT CAUSE: A ROOT CAUSE CANNOT BE EXPLICITLY DETERMINED. IT IS UNKNOWN IF THE BREAKAGE WAS A RESULT OF POOR PROSTHETIC DESIGN.

Additional Manufacturer Narrative · 0

CAPA (B)(4). MANUFACTURER REFERENCE: COMP (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE HAHN TAPERED IMPLANT FAILED. THE BONE GRADE IS NOTED AS GRADE III. THE PATIENT HAS NO MEDICAL OR DENTAL HISTORY NOTED. THE PATIENT PRESENTED ON (B)(6) 2019 FOR IMPLANT PLACEMENT ON TOOTH #5. THE PATIENT RETURNED (B)(6) 2021 NOTING THAT THE IMPLANT BROKE. UPON EXAM, THE PROVIDER CONFIRMS A BROKEN IMPLANT THAT RESULTED IN THE NEED TO REPLACE IT WITH HIOSSEN IMPLANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1119590 HAHN TAPERED IMPLANT 03.5 X11.5MM HAHN TAPERED IMPLANT DZE PRISMATIK DENTALCRAFT, INC. 70-1154-IMP0006 6051418

Patients

Seq Age Sex Outcome Treatment
1 63 YR Male Required Intervention