FDA Adverse Event Injury Summary report: N

IMRIS IMRI 3T SKYRA SYSTEM

MDR report key: 13516141 · Received February 11, 2022

Report

Report Number
3010326005-2017-00002
Event Type
Injury
Date Received
February 11, 2022
Date of Event
April 4, 2017
Report Date
April 4, 2017
Manufacturer
IMRIS - DEERFIELD IMAGING, INC
Product Code
LNH
UDI-DI
00857534006035
PMA / PMN Number
K133692
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE IS A KNOWN RISK THAT THE INTRODUCTION OF FERROUS OBJECTS INTO THE MR SUITE CAN CAUSE DAMAGE OR INJURY IF THEY ARE DRAWN TOWARDS THE MAGNET. ALL SUITES ARE EQUIPPED WITH SAFETY LABELS INFORMING USERS AND VISITORS OF THE HAZARDS INTRINSIC TO THE MR ENVIRONMENT. THE UNDERLYING CAUSE OF THE INCIDENT WAS USER ERROR. THE SERVICE TECHNICIAN FAILED TO VERIFY THAT THE MAGNET WAS DEPLOYED AND THAT THE SERVER ON THE EQUIPMENT RACK WAS LOCATED IN AN MR UNSAFE ZONE. THIS MDR IS BEING SUBMITTED OUTSIDE OF THE REQUIRED TIMEFRAME AS PART OF REMEDIAL ACTION INITIATED BY THE MANUFACTURER, IN RESPONSE TO INTERNALLY IDENTIFIED ISSUES REGARDING FAILED ELECTRONIC SUBMISSIONS THROUGH WEBTRADER.

Description of Event or Problem · 0

ON (B)(6) 2017, A SIEMENS SERVICE TECHNICIAN WAS WORKING ON A SOFTWARE UPGRADE IN THE EQUIPMENT ROOM POSITIONED ADJACENT TO THE MAGNET IN THE DIAGNOSTIC ROOM. AS THE TECHNICIAN WAS ATTEMPTING TO REMOVE AND REPLACE A COMPUTER SERVER FROM THE EQUIPMENT RACK, THE SERVER ATTRACTED TO THE MAGNET. THE TECHNICIAN'S LEFT HAND WAS CAUGHT BETWEEN THE SERVER AND THE MAGNET. THE TECHNICIAN WAS ABLE TO EXTRICATE HIS LEFT HAND FROM BETWEEN THE MAGNET AND THE EQUIPMENT, AND DAMAGE TO THE TECHNICIAN'S MIDDLE FINGER OCCURRED DURING THIS PROCESS. THE TECHNICIAN WAS TREATED IN THE EMERGENCY ROOM ONSITE. THE MAGNET WAS RAMPED DOWN TO RETRIEVE THE SERVER FROM THE SIDE OF THE MAGNET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
91339 IMRIS IMRI 3T SKYRA SYSTEM MRI MAGNET LNH IMRIS - DEERFIELD IMAGING, INC 3T SKYRA 00857534006035

Patients

Seq Age Sex Outcome Treatment
1 30 YR Male Hospitalization