FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1,11.5,MTX,MG

MDR report key: 13513589 · Received February 11, 2022

Report

Report Number
0002023141-2022-00314
Event Type
Injury
Date Received
February 11, 2022
Date of Event
April 26, 2021
Report Date
July 12, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019836
PMA / PMN Number
K101977
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TH
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

SIMILAR COMPLAINTS FOR IMPLANT INFECTION HAVE BEEN PREVIOUSLY INVESTIGATED. REFER TO ATTACHED SUMMARY INVESTIGATION. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. WHILE NON-CONFORMANCES WERE IDENTIFIED FOR SOME LOTS DURING MANUFACTURING RECORD REVIEWS, THE DOCUMENTED DISPOSITION ACTIONS FOR EACH DID NOT SUGGEST THE LIKELY RELEASE OF NON-CONFORMING PRODUCT. ZIMMER BIOMET QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. ADDITIONALLY, ALL DEVICE HISTORY RECORD REVIEWS VERIFIED THAT EACH IMPLANT WAS STERILIZED PER PROCEDURE FOR EVERY DEVICE. ALL COMPLAINT DATA USED FOR THE SUMMARY INVESTIGATION WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/IE ESCALATION. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1240117). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1240117) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION. H3 OTHER TEXT : DEVICE NOT RETURNED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DR. INDICATED INFECTION. NO OTHER IMPLANT PLACED, NO DELAY. PAIN.TOOTH SITE # 20.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
91181 IMP,TSV,4.1,11.5,MTX,MG DENTAL IMPLANT DZE ZIMMER DENTAL TSVT4B11 1240117 00889024019836

Patients

Seq Age Sex Outcome Treatment
1 39 YR Female Required Intervention