FDA Adverse Event
Malfunction
Summary report: N
TRANSTAR STRETCHER
MDR report key: 1350719
·
Received January 19, 2009
Report
- Report Number
- 1824206-2009-00276
- Event Type
- Malfunction
- Date Received
- January 19, 2009
- Date of Event
- December 17, 2008
- Report Date
- December 17, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
BRAKE NOT ENGAGING. FOUND HEX ROD OUT OF PLACE. SET SCREW BROKEN OFF IN HEX ROD. SCREW COULD NOT BE EXTRACTED. HED ROD REPLACED. BRAKES WORK FINE. TECHNICIAN STATED THAT THERE WERE NOT ANY INJURIES ASSOCIATED WITH THIS ISSUE. HE FOUND THIS STRETCH IN STORAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSTAR STRETCHER | WHEELED STRETCHER | FPO | HILL-ROM RITTER | P8000B5SSH13PN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |