670G INSULIN PUMP MMT-1780KL
Report
- Report Number
- 2032227-2022-111618
- Event Type
- Malfunction
- Date Received
- February 10, 2022
- Date of Event
- February 8, 2022
- Report Date
- June 21, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000283513
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE PUMP POWERED UP PROPERLY AFTER BATTERY INSTALLATION. THE PUMP PASSED THE DISPLACEMENT TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT AND SELF TEST. THE PUMP WAS MONITORED FOR SEVERAL HOURS AND NO BLANK DISPLAY NOTED. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. NO ALARMS/ALERTS NOTED DURING TESTING. NO POWER ERROR 25 AND REPLACE BATTERY ALERT/REPLACE BATTERY NOW ALARM RECORDED IN THE FORMATTED HISTORY FILE FOR THE EVENT DATE. HOWEVER, LOW BATTERY ALERT WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 02/08/2022 20:55:00.000 INSERT BATTERY ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 02/08/2022 21:10:50.000 02/08/2022 21:20:00.000 AND POWER LOSS ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 02/08/2022 21:21:17.000 02/08/2022 21:21:28.000 UPON CHECKING ON THE POWER DATA/DETAIL TRACE FILE, LOW BATTERY ALERT AND POWER LOSS ALARM WERE EXPECTED SINCE THE BATTERY HAS LOW/NO POWER. THE CUSTOMER MAY HAVE USED A LOW/DEPLETED BATTERY. THE PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND IT WAS VERIFIED THAT THE BATTERY TUBE POWER CONNECTOR IS PROPERLY CONNECTED TO J7/PCB1. NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A PILLOWING KEYPAD OVERLAY, A CRACKED BATTERY TUBE THREADS, A SCRATCHED CASE AND A CRACKED CASE-CORNER OF BELT CLIP RAILS NEAR THE BATTERY TUBE COMPARTMENT. BLANK DISPLAY WAS NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BLANK DISPLAY. CUSTOMER STATED THE DISPLAY WAS NOT BLANK LESS THAN 30 SECONDS AND DID NOT RETURN. CUSTOMER STATED DISPLAY WAS BLANK CURRENTLY. CUSTOMER STATED THE CONTACTS ON THE BATTERY CAP WERE NOT MISSING OR DAMAGED. CUSTOMER STATED BATTERY COMPARTMENT AND SPRING WERE NEITHER DAMAGED NOR CORRODED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND DISPLAY DID NOT RETURN BACK. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 24435 | 670G INSULIN PUMP MMT-1780KL | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KL | HG4FHC8 | 000000763000283513 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR | Female |