Report Number
3012307300-2022-03069
Event Type
Malfunction
Date Received
February 9, 2022
Report Date
March 17, 2022
Manufacturer
ST PAUL
Product Code
FRN
UDI-DI
10610586043567
PMA / PMN Number
K111386
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: ONE SMITHS MEDICAL MEDFUSION PUMP WAS RETURNED FOR ANALYSIS WITH A CRACKS ON THE TOP CASE CORNERS, BOTTOM CASE BY L-BRACKET, AND BOTH PLUNGER CASES. EVENT LOG HISTORY WAS PERFORMED AND INDICATED THAT PRIMARY AUDIBLE ALARM WAS RECORDED IN HISTORY. DURING ANALYSIS, THE REPORTED ISSUE WAS UNABLE TO BE DUPLICATED. BASED ON THE EVIDENCE, THE COMPLAINT WAS CONFIRMED, AND THE PROBLEM SOURCE OF THE REPORTED EVENT IS UNKNOWN. SERVICE REPLACED SPEAKER FOR PREVENTIVE.

Additional Manufacturer Narrative · 0

OTHER, OTHER TEXT: ADDITIONAL EVENT INFORMATION WAS RECEIVED AND B5 WAS UPDATED.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: ONE SMITHS MEDICAL MEDFUSION PUMP WAS RETURNED FOR ANALYSIS WITH A CRACKS ON THE TOP CASE CORNERS, BOTTOM CASE BY L-BRACKET, AND BOTH PLUNGER CASES. EVENT LOG HISTORY WAS PERFORMED AND INDICATED THAT PRIMARY AUDIBLE ALARM WAS RECORDED IN HISTORY. DURING ANALYSIS, THE REPORTED ISSUE WAS UNABLE TO BE DUPLICATED. BASED ON THE EVIDENCE, THE COMPLAINT WAS CONFIRMED, AND THE PROBLEM SOURCE OF THE REPORTED EVENT IS UNKNOWN. SERVICE REPLACED SPEAKER FOR PREVENTIVE.

Description of Event or Problem · 0

IT WAS REPORTED THAT CONTINUOUS PRIMARY AUDIBLE ALARM WAS EXPERIENCED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED AND ATTACHED BY ICU MEDICAL ON 11-FEB-2022:NOT INVOLVED WITH PATIENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT CONTINUOUS PRIMARY AUDIBLE ALARM WAS EXPERIENCED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
257722 MEDFUSION PUMP, INFUSION FRN ST PAUL 4000-0105-51 10610586043567

Patients

Seq Age Sex Outcome Treatment
1 Unknown