FDA Adverse Event
Malfunction
Summary report: N
MEDEX LOGICAL PRESSURE MONITORING SYSTEM
MDR report key: 13500512
·
Received February 9, 2022
Report
- Report Number
- 3012307300-2022-03059
- Event Type
- Malfunction
- Date Received
- February 9, 2022
- Date of Event
- June 4, 2021
- Report Date
- February 9, 2022
- Manufacturer
- SMITHS MEDICAL INTERNATIONAL, LTD.
- Product Code
- DRS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE WAS RETURNED TO SMITHS MEDICAL. MEASURED THE DIMENSIONS OF THE GASKETS AS RETURNED AND ALL APPEARED TO MEET SPECIFICATION. A CABLE CONNECTOR WAS CONNECTED TO EACH UNIT AND EACH UNIT SHOWED SLIGHT COMPRESSION OF THE GASKET WHEN CONNECTED AS REQUIRED. FUNCTIONALITY WAS TESTED PER INTERNAL TEST PROCEDURE OP.000272. UNABLE TO DUPLICATE THE REPORTED FAILURE. DHR: NO CAUSES OR POTENTIAL CAUSES TO THE CUSTOMER'S REPORTED PROBLEM WERE FOUND DURING THE DHR REVIEW.
Description of Event or Problem · 0
IT WAS REPORTED THAT THE CONNECTION POINT CABLE PRESSURE PICKUP NOT TIGHT. HUMIDITY ENTERS. ZERO ADJUSTMENT NO LONGER POSSIBLE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 328264 | MEDEX LOGICAL PRESSURE MONITORING SYSTEM | TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR | DRS | SMITHS MEDICAL INTERNATIONAL, LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |