FDA Adverse Event Malfunction Summary report: N

MEDEX LOGICAL PRESSURE MONITORING SYSTEM

MDR report key: 13500489 · Received February 9, 2022

Report

Report Number
3012307300-2022-03056
Event Type
Malfunction
Date Received
February 9, 2022
Date of Event
June 4, 2021
Report Date
February 9, 2022
Manufacturer
SMITHS MEDICAL INTERNATIONAL, LTD.
Product Code
DRS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO SMITHS MEDICAL. MEASURED THE DIMENSIONS OF THE GASKETS AS RETURNED AND ALL APPEARED TO MEET SPECIFICATION. A CABLE CONNECTOR WAS CONNECTED TO EACH UNIT AND EACH UNIT SHOWED SLIGHT COMPRESSION OF THE GASKET WHEN CONNECTED AS REQUIRED. FUNCTIONALITY WAS TESTED PER INTERNAL TEST PROCEDURE OP.000272. UNABLE TO DUPLICATE THE REPORTED FAILURE. DHR: NO CAUSES OR POTENTIAL CAUSES TO THE CUSTOMER'S REPORTED PROBLEM WERE FOUND DURING THE DHR REVIEW.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CONNECTION POINT CABLE PRESSURE PICKUP NOT TIGHT : HUMIDITY ENTERS : ZERO ADJUSTMENT NO LONGER POSSIBLE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
327199 MEDEX LOGICAL PRESSURE MONITORING SYSTEM TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR DRS SMITHS MEDICAL INTERNATIONAL, LTD.

Patients

Seq Age Sex Outcome Treatment
1 Unknown