FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 13492899 · Received February 9, 2022

Report

Report Number
9610877-2022-50647
Event Type
Malfunction
Date Received
February 9, 2022
Date of Event
January 19, 2022
Report Date
February 9, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDT
UDI-DI
04961333233007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL ED34-I10T2-US  IS AVAILABLE IN THE USA WITH A 510K NUMBER K192245. WE CHECKED THE RETURNED UNIT AND CONFIRMED THE CCD MODULE AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
845463 PENTAX VIDEO DUODENOSCOPE FDT HOYA CORPORATION PENTAX TOKYO OFFICE ED34-I10T2 04961333233007

Patients

Seq Age Sex Outcome Treatment
1 Unknown