FDA Adverse Event Injury Summary report: N

GENTLEPOWER LUX CONTRA ANGLE 25LPA

MDR report key: 1349193 · Received March 24, 2009

Report

Report Number
8010493-2009-00004
Event Type
Injury
Date Received
March 24, 2009
Report Date
January 11, 2008
Manufacturer
KALTENBACH & VOIGT GMBH
Product Code
EFB
PMA / PMN Number
K073478
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT WAS PRESCRIBED 800MG IBUPROFEN AND AMOXICILLIN. PATIENT WAS SEEN TWO WEEKS LATER AND WAS HEALING. DURING HANDPIECE EVALUATION, IT WAS FOUND THAT THE BEARINGS WERE WORN AND GRITTY IN THE DRIVE ASSEMBLY AND SHAFT. LOW DEBRIS WAS PRESENT. THE DOCTOR REQUESTED THAT THE HANDPIECE BE REPLACED. REPLACEMENT HANDPIECE WAS SENT.

Description of Event or Problem · 1

THE HANDPIECE HEATED UP AND A PATIENT WAS BURNED ON THE CHEEK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX CONTRA ANGLE 25LPA DENTAL HANDPIECE AND ACCESSORIES EFB KALTENBACH & VOIGT GMBH 25LPA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention