BD¿ POSIFLUSH¿ SALINE XS 10 ML
Report
- Report Number
- 9616657-2022-00003
- Event Type
- Malfunction
- Date Received
- February 7, 2022
- Date of Event
- January 21, 2022
- Report Date
- February 9, 2022
- Manufacturer
- BECTON, DICKINSON AND CO.
- Product Code
- NGT
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 306572 AND LOT NUMBER 1203643. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE NOT RETURNED FOR THIS INCIDENT, A THOROUGH SAMPLE ANALYSIS COULD NOT BE COMPLETED BY OUR QUALITY ENGINEER TEAM. BASED ON THE LIMITED INVESTIGATION RESULTS, AN EXACT CAUSE RELATED TO THE MANUFACTURING PROCESS COULD NOT BE DETERMINED FOR THIS ISSUE.
IT WAS REPORTED WHILE USING BD¿ POSIFLUSH¿ SALINE XS 10 ML THE SCALE MARKING WAS MISSING ON TWO SYRINGES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: PRODUCT: BD POSIFLUSH XS 10 ML LOT NUMBER: 1203643 PRODUCT EXPIRY DATE: 30-JUN-2024 NUMBER OF DEFECTIVE PRODUCT(S): 2 IS SAMPLE AVAILABLE: YES UNABLE TO VERIFY THAT THE FLUID IN THE SYRINGE MATCHES THE DESCRIPTION ON THE PACKAGE. SYRINGE WAS LACKING ALL APPROPRIATE MARKINGS AND LABELLING.
IT WAS REPORTED WHILE USING BD¿ POSIFLUSH¿ SALINE XS 10 ML THE SCALE MARKING WAS MISSING ON TWO SYRINGES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: PRODUCT: BD POSIFLUSH XS 10 ML. LOT NUMBER: 1203643. PRODUCT EXPIRY DATE: 30-JUN-2024. NUMBER OF DEFECTIVE PRODUCT(S): 2. IS SAMPLE AVAILABLE: YES. UNABLE TO VERIFY THAT THE FLUID IN THE SYRINGE MATCHES THE DESCRIPTION ON THE PACKAGE. SYRINGE WAS LACKING ALL APPROPRIATE MARKINGS AND LABELLING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 501471 | BD¿ POSIFLUSH¿ SALINE XS 10 ML | SALINE, VASCULAR ACCESS FLUSH | NGT | BECTON, DICKINSON AND CO. | 1203643 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |