FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 13468731 · Received February 7, 2022

Report

Report Number
9610877-2022-50415
Event Type
Malfunction
Date Received
February 7, 2022
Date of Event
January 18, 2022
Report Date
February 7, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG-1690K IS AVAILABLE IN THE USA WITH A 510K NUMBER K131902. WE CHECKED THE RETURNED UNIT AND CONFIRMED THE CCD DRIVER PCB AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD DRIVER PCB. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
488040 PENTAX VIDEO GASTROSCOPE FDS HOYA CORPORATION PENTAX TOKYO OFFICE EG16-K10

Patients

Seq Age Sex Outcome Treatment
1 Unknown