FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 13454136 · Received February 4, 2022

Report

Report Number
1221359-2022-00738
Event Type
Malfunction
Date Received
February 4, 2022
Date of Event
December 28, 2021
Report Date
February 15, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M164094 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000J/ LOT: M164094, TEST BASE PART NUMBER 190-000J/ LOT: M164094. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M164094 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION. A REVIEW OF COMPLAINTS AGAINST THE KIT LOT FOR THE REPORTED ISSUE INDICATES PRODUCT PERFORMANCE IS AS EXPECTED AND A PRODUCT DEFICIENCY HAS NOT BEEN IDENTIFIED. RELATED REPORT: 1221359-2022-00739.

Additional Manufacturer Narrative · 0

THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE PROVIDED. RELATED REPORT: 1221359-2022-00739.

Additional Manufacturer Narrative · 0

THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE PROVIDED. RELATED REPORT: 1221359-2022-00739.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TWO FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY USING UNSPECIFIED SAMPLE TYPES. THE FIRST COVID-19 ASSAY WAS PERFORMED ON (B)(6) 2021 WHICH GENERATED A POSITIVE RESULT. THE SECOND COVID-19 ASSAY WAS PERFORMED ON (B)(6) 2021 WHICH GENERATED A POSITIVE RESULT. CONFIRMATION PCR TESTING WAS PERFORMED FOR BOTH PATIENTS ON UNKNOWN DATES, USING UNSPECIFIED SAMPLE TYPES, WHICH GENERATED NEGATIVE RESULTS. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED. THIS MANUFACTURER REPORT ADDRESSES FALSE POSITIVE RESULT 1 OF 2.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TWO FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY USING UNSPECIFIED SAMPLE TYPES. THE FIRST COVID-19 ASSAY WAS PERFORMED ON (B)(6) 2021 WHICH GENERATED A POSITIVE RESULT. THE SECOND COVID-19 ASSAY WAS PERFORMED ON (B)(6) 2021 WHICH GENERATED A POSITIVE RESULT. CONFIRMATION PCR TESTING WAS PERFORMED FOR BOTH PATIENTS ON UNKNOWN DATES, USING UNSPECIFIED SAMPLE TYPES, WHICH GENERATED NEGATIVE RESULTS. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED. THIS MANUFACTURER REPORT ADDRESSES FALSE POSITIVE RESULT 1 OF 2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1649477 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M164094

Patients

Seq Age Sex Outcome Treatment
1 Unknown