BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE
Report
- Report Number
- 3003152976-2022-00046
- Event Type
- Malfunction
- Date Received
- February 4, 2022
- Date of Event
- January 21, 2022
- Report Date
- March 9, 2022
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED AS A DEFAULT A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED AS A DEFAULT A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY: TWELVE SAMPLES AND ONE PHOTO WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION OF THE PHOTO, SILICONE CAN BE OBSERVED, HOWEVER, IT CANNOT BE CONFIRMED IT IS IN EXCESS. ALL PHYSICAL SAMPLES WERE EVALUATED, NO EXCESS OF SILICONE OR OTHER FOREIGN PARTICLES CAN BE OBSERVED IN ANY OF THE SAMPLES. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2110137, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE.
IT WAS REPORTED THAT BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE CONTAINED FOREIGN MATTER IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ON A BATCH OF THE BD PLASTIPAK 50 ML BATCH 2110137, IT IS NOTICEABLE THAT THERE IS AN EXCESS OF SILICONE OIL ON THE PLUNGER."
IT WAS REPORTED THAT BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE CONTAINED FOREIGN MATTER IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ON A BATCH OF THE BD PLASTIPAK 50 ML BATCH 2110137, IT IS NOTICEABLE THAT THERE IS AN EXCESS OF SILICONE OIL ON THE PLUNGER."
IT WAS REPORTED THAT BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE CONTAINED FOREIGN MATTER IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ON A BATCH OF THE BD PLASTIPAK 50 ML BATCH 2110137, IT IS NOTICEABLE THAT THERE IS AN EXCESS OF SILICONE OIL ON THE PLUNGER."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1421426 | BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2110137 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |