Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED AS A DEFAULT A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED AS A DEFAULT A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: TWELVE SAMPLES AND ONE PHOTO WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION OF THE PHOTO, SILICONE CAN BE OBSERVED, HOWEVER, IT CANNOT BE CONFIRMED IT IS IN EXCESS. ALL PHYSICAL SAMPLES WERE EVALUATED, NO EXCESS OF SILICONE OR OTHER FOREIGN PARTICLES CAN BE OBSERVED IN ANY OF THE SAMPLES. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2110137, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE CONTAINED FOREIGN MATTER IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ON A BATCH OF THE BD PLASTIPAK 50 ML BATCH 2110137, IT IS NOTICEABLE THAT THERE IS AN EXCESS OF SILICONE OIL ON THE PLUNGER."

Description of Event or Problem · 0

IT WAS REPORTED THAT BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE CONTAINED FOREIGN MATTER IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ON A BATCH OF THE BD PLASTIPAK 50 ML BATCH 2110137, IT IS NOTICEABLE THAT THERE IS AN EXCESS OF SILICONE OIL ON THE PLUNGER."

Description of Event or Problem · 0

IT WAS REPORTED THAT BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE CONTAINED FOREIGN MATTER IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ON A BATCH OF THE BD PLASTIPAK 50 ML BATCH 2110137, IT IS NOTICEABLE THAT THERE IS AN EXCESS OF SILICONE OIL ON THE PLUNGER."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1421426 BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2110137

Patients

Seq Age Sex Outcome Treatment
1 Unknown