FDA Adverse Event Injury Summary report: N

SIMPLE PREP CARBIDE

MDR report key: 13446441 · Received February 3, 2022

Report

Report Number
2416455-2022-00001
Event Type
Injury
Date Received
February 3, 2022
Date of Event
January 1, 2022
Report Date
April 21, 2022
Manufacturer
COLTENE WHALEDENT INC.
Product Code
EJL
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: (B)(4) UNITS OF THIS KIT WERE PUT TO STOCK IN SEPTEMBER 2021. THERE ARE NO UNITS REMAINING. COMPLAINT ANALYSIS FOR THE LAST 12 MONTHS IDENTIFIES NO ADDITIONAL COMPLAINTS FOR THIS PRODUCT. NO PRODUCT HAS BEEN RETURNED FOR INVESTIGATION. THESE KITS WERE PACKAGED WITH SEMI-FINISHED BURS 50008508 K91987. THIS BATCH RECORD WAS REVIEWED. ALL DIMENSIONAL REQUIREMENTS AND THE FUNCTIONAL TEST FOR NECK STRENGTH WERE MET. THE CUSTOMER COMPLAINT CANNOT BE CONFIRMED. ROOT CAUSE: NO TRUE ROOT CAUSE COULD BE IDENTIFIED IN THIS COMPLAINT. POSSIBLE ROOT CAUSES INCLUDE: THE BURS MAY HAVE BEEN DAMAGED UPON ARRIVAL TO THE CUSTOMER BUT THE CUSTOMER MAY NOT HAVE NOTICED PRIOR TO USING THE PRODUCT. THE BURS MAY HAVE BEEN USED MULTIPLE TIMES AND THE FLUTE EDGES MAY HAVE BEEN WORN. USING A BUR WITH EXCESSIVE WEAR MAY CAUSE THE BUR TO BREAK OR CAUSE INJURY. COLD DISINFECTANT SOLUTIONS MAY HAVE BEEN USED THAT CONTAIN CORROSIVE AGENTS THAT DEGRADE THE PERFORMANCE AND STRENGTH OF THE CARBIDE BURS. THE BURS MAY NOT HAVE BEEN STORED IN A MOISTURE FREE ENVIRONMENT. THE BURS MAY HAVE BEEN USED WITH HEAVY TRANSVERSE CUTTING PRESSURE.

Description of Event or Problem · 0

BUR HEAD BREAKS OFF. BUR BROKE WHILE USING ON PATIENT. DOCTOR COULD NOT FIND BROKEN TIP PATIENT WAS SENT TO HAVE XRAYS TO MAKE SURE TIP WAS NOT SWALLOWED. NO SURGICAL INTERVENTION WAS REQUIRED. ON (B)(6) 2022 ADDITIONAL INFORMATION WAS PROVIDED: FEMALE (B)(6). CHEST X-RAY FOUND NOTHING. DOCTOR SAID THAT HE THINKS THEY FOUND THE BROKEN TIP BUT WAS NOT FOR SURE. THIS IS WHY THEY DID CHEST X-RAY FOR PATIENT SAFETY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1451216 SIMPLE PREP CARBIDE CARBIDE BUR EJL COLTENE WHALEDENT INC. K99739

Patients

Seq Age Sex Outcome Treatment
1 35 YR Female Required Intervention